FDA Adverse Event Malfunction Summary report: N

LC4200 PCA AUST

MDR report key: 1883198 · Received October 20, 2010

Report

Report Number
2921482-2010-00804
Event Type
Malfunction
Date Received
October 20, 2010
Date of Event
September 21, 2010
Report Date
September 22, 2010
Manufacturer
HOSPIRA, INC.
Product Code
MEA
PMA / PMN Number
UNK
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

TESTING AND INVESTIGATION FOUND THE DEVICE DID NOT ALARM WHEN A DISTAL OCCLUSION WAS PRESENT. THIS WAS DUE TO WORN THREADS ON THE HALF NUT. THIS ALLOWED THE HALF NUT TO SLIP ON THE LEAD SCREW WHEN OCCLUSION PRESSURE BUILT. THE CATALOG NUMBER PROVIDED IS THE INTERNATIONAL LIST NUMBER THAT WAS INVOLVED IN THE REPORTABLE EVENT, WHICH IS COMPARABLE TO THE DOMESTIC LIST NUMBER LISTED IN THE "OTHER" FIELD. THE DOMESTIC LIST NUMBER HAS A COMMON DEVICE NAME OF 80FRN AND HAS A 510K OF K912928. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THAT DURING PREVENTIVE MAINTENANCE TESTING AT THE USER FACILITY, THE DEVICE DID NOT ALARM WHEN A DISTAL OCCLUSION WAS PRESENT. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LC4200 PCA AUST UNK MEA HOSPIRA, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA