FDA Adverse Event
Malfunction
Summary report: N
INOMAX DS (DELIVERY SYSTEM)
MDR report key: 1883187
·
Received October 20, 2010
Report
- Report Number
- 3004531588-2010-00105
- Event Type
- Malfunction
- Date Received
- October 20, 2010
- Date of Event
- May 17, 2010
- Report Date
- October 20, 2010
- Manufacturer
- INO THERAPEUTICS LLC/IKARIA
- Product Code
- MRN
- PMA / PMN Number
- K061901
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
EVAL SUMMARY: THE INVESTIGATION OF THE DEVICE IS COMPLETE AND IS AS FOLLOWS: THE DEVICE WAS CONFIRMED TO HAVE A LEAK THAT WAS TRACED TO THE PRESSURE SWITCH. THE PRESSURE SWITCH WAS REPLACED AND IT WAS CONFIRMED THE LEAK STOPPED AND WAS CORRECTED. THE ROOT CAUSE OF THE INCIDENT WAS A FAILED PRESSURE SWITCH.
Description of Event or Problem · 1
ON (B)(6), 2010, A RESPIRATORY THERAPIST REPORTED LEAKING FROM THE BACK OF A INOMAX DS, #(B)(4). THE DEVICE WAS NOT ON A PT AND NO ADVERSE EVENT WAS REPORTED BY THE RESPIRATORY THERAPIST. THE DEVICE WAS REMOVED FROM SERVICE BY THE CUSTOMER AND RETURNED TO THE COMPANY FOR INVESTIGATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INOMAX DS (DELIVERY SYSTEM) | APPARATUS, NITRIC OXIDE DELIVERY | MRN | INO THERAPEUTICS LLC/IKARIA | 10003 |