VASOVIEW HEMOPRO EVH SYSTEM
Report
- Report Number
- 2242352-2010-03137
- Event Type
- Malfunction
- Date Received
- October 19, 2010
- Date of Event
- September 17, 2010
- Report Date
- September 20, 2010
- Manufacturer
- MAQUET CARDIOVASCULAR, LLC
- Product Code
- GEI
- PMA / PMN Number
- K052274
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
INVESTIGATION SUMMARY: THE DEVICE WAS RETURNED TO MAQUET CARDIAC SURGERY ON 10/04/2010, FOR INVESTIGATION. VISUAL INSPECTION REVEALED THAT THERE WERE NO NON-CONFORMITIES WITH THE DEVICE. THERE WAS SOME EVIDENCE OF BLOOD. A PRE-CAUTERY TEST WAS PERFORMED ON THE DEVICE WITH A REFERENCE POWER SUPPLY AND EXTENSION CABLE. THE DEVICE PASSED THE PRE-CAUTERY TEST; IT PRODUCED ENERGY AND STEAM, AND DID NOT REMAIN ACTIVATED. BASED UPON THESE FINDINGS, THE REPORTED FAILURE, "DEVICE OVERHEATED" COULD NOT BE CONFIRMED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE PRODUCT. THERE WAS NO NON-CONFORMANCE WHICH COULD HAVE CONTRIBUTED TO THIS FAILURE MODE. (B)(4).
THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, THE VASOVIEW HEMOPRO ENDOSCOPIC VESSEL HARVESTING SYSTEM WOULD NOT TURN OFF. A REPLACEMENT UNIT WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PATIENT EFFECTS. THE PRODUCT WAS RETURNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VASOVIEW HEMOPRO EVH SYSTEM | ENDOSCOPIC VESSEL HARVESTING SYSTEM | GEI | MAQUET CARDIOVASCULAR, LLC | VH-3000 | 25017307 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |