FDA Adverse Event Malfunction Summary report: N

PUMP MMT-522NAS PRDGM INS V2.2 SK EN

MDR report key: 1882981 · Received October 22, 2010

Report

Report Number
3004209178-2010-83187
Event Type
Malfunction
Date Received
October 22, 2010
Date of Event
October 12, 2010
Report Date
October 12, 2010
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
MDS
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR. TROUBLESHOOTING WAS PERFORMED. THE CUSTOMER STATED THAT THE DEVICE WAS STUCK IN REWIND MODE, AND THEN SHE WAS ABLE TO EXIT FROM REWIND MODE. RAN A DISPLACEMENT TEST AND THE INSULIN PUMP FAILED THE TEST. NO FURTHER INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUMP MMT-522NAS PRDGM INS V2.2 SK EN INFUSION PUMP MDS MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-522NAS

Patients

Seq Age Sex Outcome Treatment
1 39 YR