FDA Adverse Event
Malfunction
Summary report: N
CAREASSIST BED
MDR report key: 1882960
·
Received October 22, 2010
Report
- Report Number
- 9615739-2010-00239
- Event Type
- Malfunction
- Date Received
- October 22, 2010
- Date of Event
- September 28, 2010
- Report Date
- September 28, 2010
- Manufacturer
- HILL-ROM PLUVIGNER
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE TECHNICIAN FOUND ALL FOUR CASTERS HAD BROKEN BRAKE STEMS. THE TECHNICIAN INDICATED THE CASTERS WERE ENGAGED TOO HARD PAST THE POINT OF LOCKING. THE TECHNICIAN REPLACED ALL FOUR CASTERS TO REPAIR THE BED.
Description of Event or Problem · 1
INFO REC'D INDICATES THE BRAKE WILL NOT HOLD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAREASSIST BED | AC POWERED HOSPITAL BED | FNL | HILL-ROM PLUVIGNER | 1170 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |