FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN INFUSION PUMP
MDR report key: 1882924
·
Received October 22, 2010
Report
- Report Number
- 2531779-2010-01862
- Event Type
- Malfunction
- Date Received
- October 22, 2010
- Date of Event
- September 22, 2010
- Report Date
- September 23, 2010
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SW
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE PUMP HAS BEEN RETURNED TO ANIMAS FOR EVALUATION. EVALUATION REVEALED THAT THE PUMP WAS INTERMITTENTLY UNRESPONSIVE TO OK BUTTON, AND UP AND DOWN ARROW BUTTON PRESSES. THE KEYPAD WAS REMOVED AND ADHESIVE/CONTAMINATION WAS OBSERVED UNDER THE BUTTON CONTACTS. THE KEYPAD WAS TORN NEAR THE UP ARROW BUTTON.
Description of Event or Problem · 1
THE DISTRIBUTOR CONTACTED ANIMAS ALLEGING THAT THE PUMP WAS HAVING DIFFICULTY RESPONDING TO BUTTON PRESSES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ANIMAS 2020 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |