FDA Adverse Event
Malfunction
Summary report: N
ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM
MDR report key: 1882869
·
Received October 22, 2010
Report
- Report Number
- 2531779-2010-01792
- Event Type
- Malfunction
- Date Received
- October 22, 2010
- Date of Event
- September 21, 2010
- Report Date
- September 21, 2010
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE KEYPAD BUTTONS WERE FOUND TO BE INTERMITTENTLY RESPONDING TO PRESSES. THE KEYPAD WAS REMOVED AND ADHESIVE WAS OBSERVED UNDER THE BUTTON CONTACTS.
Description of Event or Problem · 1
THE PATIENT REPORTED THAT THE KEYPAD IS NOT RESPONDING, THE 'UP' AND 'DOWN' BUTTONS HAVE BEEN INTERMITTENTLY UNRESPONSIVE FOR THE PAST WEEK. THE PATIENT REPORTS NOT EXPOSING THE PUMP TO MOISTURE AND THE KEYPAD IS INTACT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ONE TOUCH PING | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 44 YR |