FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1882869 · Received October 22, 2010

Report

Report Number
2531779-2010-01792
Event Type
Malfunction
Date Received
October 22, 2010
Date of Event
September 21, 2010
Report Date
September 21, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE KEYPAD BUTTONS WERE FOUND TO BE INTERMITTENTLY RESPONDING TO PRESSES. THE KEYPAD WAS REMOVED AND ADHESIVE WAS OBSERVED UNDER THE BUTTON CONTACTS.

Description of Event or Problem · 1

THE PATIENT REPORTED THAT THE KEYPAD IS NOT RESPONDING, THE 'UP' AND 'DOWN' BUTTONS HAVE BEEN INTERMITTENTLY UNRESPONSIVE FOR THE PAST WEEK. THE PATIENT REPORTS NOT EXPOSING THE PUMP TO MOISTURE AND THE KEYPAD IS INTACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1 44 YR