FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1882855 · Received October 22, 2010

Report

Report Number
2531779-2010-01810
Event Type
Malfunction
Date Received
October 22, 2010
Date of Event
September 21, 2010
Report Date
September 21, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS. THE EVALUATION REVEALED THAT THE CONTRAST AND DOWN BUTTONS WERE INTERMITTENTLY RESPONDING. THE DOWN ARROW BUTTON CONTACT WAS ALSO MISALIGNED. IN ADDITION, THE AUDIO BOLUS, UP AND OK BUTTONS WERE RESPONSIVE AND SCROLLING WAS NOT OBSERVED DURING INVESTIGATION.

Description of Event or Problem · 1

THE PT REPORTED KEYPAD RESPONSES ARE MORE SENSITIVE AND RESPOND TO THE SLIGHTEST TOUCH. THE PT ALSO REPORTED THAT WHEN USING THE AUDIO BOLUS BUTTON DELIVERY IS INTERMITTENTLY CANCELLING DUE TO BUTTON PRESSES. THE 'UP' AND 'DOWN' BUTTONS RESPOND TO THE SLIGHTEST PRESS AND SCROLLS THROUGH SCREENS TOO QUICKLY AND BUTTONS ARE STICKING. THE RPTR DENIES THAT THE PUMP IS EXPOSED TO MOISTURE OR CLEANING SOLVENTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1 57 YR