FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1882850 · Received October 22, 2010

Report

Report Number
2531779-2010-01796
Event Type
Malfunction
Date Received
October 22, 2010
Date of Event
September 21, 2010
Report Date
September 21, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. ALL KEYPAD BUTTONS WERE FOUND TO BE INTERMITTENTLY RESPONDING TO BUTTON PRESSES. THE KEYPAD WAS REMOVED AND ADHESIVE WAS OBSERVED UNDER THE BUTTON CONTACTS.

Description of Event or Problem · 1

THE PT REPORTED THAT FOR THE PAST 2 TO 3 WEEKS THE 'OK' 'UP' AND 'DOWN' BUTTONS HAVE BECOME INTERMITTENTLY UNRESPONSIVE. NO CHANGE IN ACTIVITY AND THE PT DENIES EXPOSING THE PUMP TO MOISTURE OR CLEANING PRODUCTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1 55 YR