FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1882840 · Received October 22, 2010

Report

Report Number
2531779-2010-01859
Event Type
Malfunction
Date Received
October 22, 2010
Report Date
September 22, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS; HOWEVER, THE INVESTIGATION HAS NOT BEEN COMPLETED. NO CONCLUSIONS CAN BE DRAWN AT THIS TIME. ONCE THE EVALUATION HAS BEEN COMPLETED A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

THE PT INDICATED AFTER HE WENT SWIMMING, THE PUMP REBOOTED AND NOW THE PUMP IS NO LONGER RESPONDING TO THE UP ARROW, DOWN ARROW, AND OK BUTTONS. THE KEYPAD IS INTACT AND NOT PEELING. WHILE TROUBLESHOOTING WITH THE ANIMAS REPRESENTATIVE, THE PT FOUND MOISTURE INSIDE THE PUMP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1