FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 1882767 · Received October 22, 2010

Report

Report Number
2531779-2010-01832
Event Type
Malfunction
Date Received
October 22, 2010
Date of Event
September 22, 2010
Report Date
September 22, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
EZ
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS FOR EVALUATION. EVALUATION REVEALED THAT THE PUMP WAS INTERMITTENTLY UNRESPONSIVE TO CONTRAST AND DOWN ARROW BUTTON PRESSES. THE KEYPAD WAS REMOVED AND ADHESIVE/CONTAMINATION WAS OBSERVED UNDER THE BUTTON CONTACTS. THE CONTRAST AND DOWN ARROW BUTTON CONTACTS WERE ALSO FOUND TO BE MISALIGNED.

Description of Event or Problem · 1

THE DISTRIBUTOR CONTACTED ANIMAS ALLEGING THAT THE PUMP WAS REACTING BADLY TO BUTTON PRESSES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORP. ANIMAS 2020 NA

Patients

Seq Age Sex Outcome Treatment
1