FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 1882720 · Received October 21, 2010

Report

Report Number
2531779-2010-01815
Event Type
Malfunction
Date Received
October 21, 2010
Date of Event
September 22, 2010
Report Date
September 22, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE PUMP WAS INTERMITTENTLY RESPONDING TO THE DOWN ARROW BUTTON AND THERE WAS ADHESIVE/CONTAMINATION FOUND UNDER THE BUTTON CONTACTS.

Description of Event or Problem · 1

THE PT CLAIMED THAT THE PUMP IS NOT RESPONDING TO THE DOWN ARROW BUTTON. THE PUMP REPORTEDLY HAS NOT BEEN EXPOSED TO MOISTURE AND THE RUBBER KEYPAD IS INTACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORP. ANIMAS 2020 NA

Patients

Seq Age Sex Outcome Treatment
1 38 YR