FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 1882707 · Received October 21, 2010

Report

Report Number
2531779-2010-01805
Event Type
Malfunction
Date Received
October 21, 2010
Date of Event
September 22, 2010
Report Date
September 22, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS FOR EVAL. EVAL REVEALED THAT THE PUMP WAS INTERMITTENTLY UNRESPONSIVE TO CONTRAST, UP ARROW, AND DOWN ARROW BUTTON PRESSES. THE OK BUTTON WAS ALSO FOUND TO BE STICKING AND RESPONDED TO HARD PRESSES. THE KEYPAD WAS REMOVED AND ADHESIVE/CONTAMINATION WAS OBSERVED UNDER THE BUTTON CONTACTS. THE KEYPAD WAS FOUND TO BE LIFTED NEAR THE UP ARROW BUTTON.

Description of Event or Problem · 1

THE RPTR CONTACTED ANIMAS ON BEHALF OF HIS SO (THE PT) ALLEGING THAT THE OK BUTTON ON THE PUMP WAS STICKING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORP. ANIMAS 2020 NA

Patients

Seq Age Sex Outcome Treatment
1 3 YR