FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN INFUSION PUMP
MDR report key: 1882707
·
Received October 21, 2010
Report
- Report Number
- 2531779-2010-01805
- Event Type
- Malfunction
- Date Received
- October 21, 2010
- Date of Event
- September 22, 2010
- Report Date
- September 22, 2010
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE PUMP HAS BEEN RETURNED TO ANIMAS FOR EVAL. EVAL REVEALED THAT THE PUMP WAS INTERMITTENTLY UNRESPONSIVE TO CONTRAST, UP ARROW, AND DOWN ARROW BUTTON PRESSES. THE OK BUTTON WAS ALSO FOUND TO BE STICKING AND RESPONDED TO HARD PRESSES. THE KEYPAD WAS REMOVED AND ADHESIVE/CONTAMINATION WAS OBSERVED UNDER THE BUTTON CONTACTS. THE KEYPAD WAS FOUND TO BE LIFTED NEAR THE UP ARROW BUTTON.
Description of Event or Problem · 1
THE RPTR CONTACTED ANIMAS ON BEHALF OF HIS SO (THE PT) ALLEGING THAT THE OK BUTTON ON THE PUMP WAS STICKING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ANIMAS 2020 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 3 YR |