ADULT DUAL-HEATED BREATHING CIRCUIT
Report
- Report Number
- 9611451-2010-00619
- Event Type
- Malfunction
- Date Received
- October 18, 2010
- Date of Event
- September 21, 2010
- Report Date
- September 21, 2010
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD.
- Product Code
- BZE
- PMA / PMN Number
- K983112
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE RT240 ADULT DUAL-HEATED BREATHING CIRCUIT IS EN ROUTE TO FISHER & PAYKEL HEALTHCARE FOR INVESTIGATION. WE WILL PROVIDE A FOLLOW-UP REPORT ONCE WE RECEIVE THE COMPLAINT DEVICE AND HAVE COMPLETED OUR INVESTIGATION.
(B)(4). METHOD: A BOX WITH AN RT340 PRODUCT LABEL BUT CONTAINING 10 RT240 BREATHING CIRCUIT KITS WAS RETURNED TO FISHER & PAYKEL HEALTHCARE (FPH) IN (B)(4) FOR VISUAL INSPECTION. RESULTS: INSPECTION CONFIRMED THE REPORTED FAULT. THE INCORRECT PRODUCT LABEL (RT340 INSTEAD OF RT240) WAS ATTACHED TO THE BOX. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 100819. CONCLUSION: THE PACKAGING HAD BEEN PRE-LABELLED WITH AN RT340 PRODUCT LABEL, HOWEVER, THE MANUFACTURING LINE CHANGED OVER TO THE RT240 BREATHING CIRCUIT. THIS RESULTED IN THE RT240 BREATHING CIRCUIT KITS BEING PACKED IN A BOX LABELED WITH RT340. THERE ARE STANDARD OPERATING PROCEDURES (SOPS) IN PLACE TO ASSIST OPERATORS ON THE PRODUCTION LINE CORRECTLY PACK BREATHING CIRCUITS. THIS INCLUDES INSTRUCTIONS STATING THAT A LABEL SHOULD ONLY BE AFFIXED TO A BOX IF THE BOX IS GOING TO BE PACKED. (B)(4).
A DISTRIBUTOR IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE'S (FPH) CUSTOMER CARE SPECIALIST IN OUR US OFFICE "THAT A CASE OF RT340 WAS MISLABELED AND INSIDE WAS RT240". THIS WAS NOTICED PRIOR TO PATIENT USE. NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADULT DUAL-HEATED BREATHING CIRCUIT | BZE | FISHER & PAYKEL HEALTHCARE LTD. | T240 | 100917 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |