FDA Adverse Event Malfunction Summary report: N

AMPLATZER TALISMAN PFO OCCLUDER

MDR report key: 18826668 · Received March 4, 2024

Report

Report Number
2135147-2024-00954
Event Type
Malfunction
Date Received
March 4, 2024
Date of Event
February 8, 2024
Report Date
April 29, 2024
Manufacturer
ABBOTT MEDICAL
Product Code
MLV
UDI-DI
05415067033314
PMA / PMN Number
P120021
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AN EVENT OF AIR WHILE LOADING THE OCCLUDER INTO THE DELIVERY SHEATH WAS REPORTED. A RETURNED DEVICE INSPECTION, TO RULE OUT ANY DEVICE-RELATED CAUSES, COULD NOT BE PERFORMED AS THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE DEVICE HISTORY RECORD WAS REVIEWED TO ENSURE THAT EACH MANUFACTURING AND INSPECTION OPERATION WAS PERFORMED, AND THE PRODUCT MET ALL SPECIFICATIONS. BASED ON THE INFORMATION RECEIVED, THE CAUSE OF THE REPORTED INCIDENT COULD NOT BE CONCLUSIVELY DETERMINED. THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO LABELING DESIGN OR MANUFACTURING OF THE DEVICE

Additional Manufacturer Narrative · 0

INVESTIGATION IS NOT YET COMPLETE. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT ON (B)(6) 2024, A 18-25MM AMPLATZER TALISMAN PATENT FORAMEN OVALE (PFO) OCCLUDER (LOT: 8864459) WAS CHOSEN FOR IMPLANTATION UTILIZING AN 8F AMPLATZER TALISMAN DELIVERY SHEATH. DURING PREPARATION, WHEN ATTEMPTING TO ADVANCE THE OCCLUDER INTO THE DELIVERY SHEATH, AIR BUBBLES APPEARED IN THE LOADER. ONLY WHEN ALL CONNECTIONS, FROM THE SHEATH TO THE LOADER, FROM THE LOADER TO THE HEMOSTASIS VALVE AND THE HEMOSTASIS VALVE WERE SUBMERGED IN A WATER BATH, COULD THE DEVICE BE ADVANCED WITHOUT BUBBLES. THE DEVICE WAS IMPLANTED SUCCESSFULLY. NO AIR ENTERED THE ANATOMY AND NO ADDITIONAL ASPIRATION WAS NEEDED TO PREVENT AIR FROM ENTERED THE PATIENT. THERE WERE NO REPORTED PATIENT CONSEQUENCES OR CLINICALLY SIGNIFICANT DELAY. THE PATIENT REMAINED HEMODYNAMICALLY STABLE THROUGHOUT THE PROCEDURE.

Description of Event or Problem · 0

IT WAS REPORTED THAT ON (B)(6) 2024, A 18-25MM AMPLATZER TALISMAN PATENT FORAMEN OVALE (PFO) OCCLUDER (LOT: 8864459) WAS CHOSEN FOR IMPLANTATION UTILIZING AN 8F AMPLATZER TALISMAN DELIVERY SHEATH. DURING PREPARATION, WHEN ATTEMPTING TO ADVANCE THE OCCLUDER INTO THE DELIVERY SHEATH, AIR BUBBLES APPEARED IN THE LOADER. ONLY WHEN ALL CONNECTIONS, FROM THE SHEATH TO THE LOADER, FROM THE LOADER TO THE HEMOSTASIS VALVE AND THE HEMOSTASIS VALVE WERE SUBMERGED IN A WATER BATH, COULD THE DEVICE BE ADVANCED WITHOUT BUBBLES. THE DEVICE WAS IMPLANTED SUCCESSFULLY. NO AIR ENTERED THE ANATOMY AND NO ADDITIONAL ASPIRATION WAS NEEDED TO PREVENT AIR FROM ENTERED THE PATIENT. THERE WERE NO REPORTED PATIENT CONSEQUENCES OR CLINICALLY SIGNIFICANT DELAY. THE PATIENT REMAINED HEMODYNAMICALLY STABLE THROUGHOUT THE PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
710552 AMPLATZER TALISMAN PFO OCCLUDER TRANSCATHETER SEPTAL OCCLUDER MLV ABBOTT MEDICAL 8864459 05415067033314

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown AMPLATZER TALISMAN DELIVERY SHEATH.