FDA Adverse Event
Malfunction
Summary report: N
AFFINITY 3
MDR report key: 1882664
·
Received October 21, 2010
Report
- Report Number
- 1824206-2010-10601
- Event Type
- Malfunction
- Date Received
- October 21, 2010
- Date of Event
- September 23, 2010
- Report Date
- September 23, 2010
- Manufacturer
- HILL-ROM, INC.
- Product Code
- HDD
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE TECHNICIAN FOUND THE CAM PIVOT LINKING THE BRAKE BAR TO THE PEDAL SNAPPED AT THE FIXING POINT. THE TECHNICIAN REPLACED THE CAM PIVOT TO REPAIR THE BED.
Description of Event or Problem · 1
INFORMATION RECEIVED INDICATES ONE OF THE TWO BRAKE CASTERS IS NOT HOLDING WHEN IN BRAKE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AFFINITY 3 | BIRTHING BED | HDD | HILL-ROM, INC. | 3700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |