FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN INFUSION PUMP
MDR report key: 1882496
·
Received October 21, 2010
Report
- Report Number
- 2531779-2010-01760
- Event Type
- Malfunction
- Date Received
- October 21, 2010
- Date of Event
- September 20, 2010
- Report Date
- September 20, 2010
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Removal / Correction Number
- 2531779-03/24/2010-003-R
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. ANIMAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. A REVIEW OF THE PUMP HISTORY INDICATED THAT LOSS OF CARTRIDGE DETECTION HAD OCCURRED. THE LOSS OF DETECTION WAS DUPLICATED DURING TESTING. INVESTIGATION REVEALED A DISLODGED DISPLAY SCREEN AND FULLY DISLODGED FORCE SENSOR PINS.
Description of Event or Problem · 1
THE PT REPORTED THAT THE PUMP DISPENSED INSULIN DURING THE LOAD CARTRIDGE PHASE. DURING THE MONTH PRECEDING THIS EVENT, THE PT HAD RESPONDED TO MULTIPLE LOSS OF PRIME WARNINGS. THE PT CONFIRMED, HE WAS DISCONNECTED FROM THE INFUSION SITE DURING THESE OCCURRENCES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ANIMAS 2020 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 30 YR |