FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 1882496 · Received October 21, 2010

Report

Report Number
2531779-2010-01760
Event Type
Malfunction
Date Received
October 21, 2010
Date of Event
September 20, 2010
Report Date
September 20, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K042873
Removal / Correction Number
2531779-03/24/2010-003-R
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. ANIMAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. A REVIEW OF THE PUMP HISTORY INDICATED THAT LOSS OF CARTRIDGE DETECTION HAD OCCURRED. THE LOSS OF DETECTION WAS DUPLICATED DURING TESTING. INVESTIGATION REVEALED A DISLODGED DISPLAY SCREEN AND FULLY DISLODGED FORCE SENSOR PINS.

Description of Event or Problem · 1

THE PT REPORTED THAT THE PUMP DISPENSED INSULIN DURING THE LOAD CARTRIDGE PHASE. DURING THE MONTH PRECEDING THIS EVENT, THE PT HAD RESPONDED TO MULTIPLE LOSS OF PRIME WARNINGS. THE PT CONFIRMED, HE WAS DISCONNECTED FROM THE INFUSION SITE DURING THESE OCCURRENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORP. ANIMAS 2020 NA

Patients

Seq Age Sex Outcome Treatment
1 30 YR