ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM
Report
- Report Number
- 2531779-2010-01773
- Event Type
- Malfunction
- Date Received
- October 21, 2010
- Date of Event
- September 22, 2010
- Report Date
- September 22, 2010
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- UNKNOWN
Narratives
THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. THE DISPLAY SCREEN WAS FOUND TO BE UNRESPONSIVE. THE KEYPAD WAS FOUND TO BE LOOSE AND PEELING. THE PUMP WAS OPENED AND MOISTURE AND CORROSION WERE OBSERVED ON CIRCUIT BOARD ON THE DISPLAY AND THE KEYPAD FLEX CONNECTORS. A REVIEW OF THE PUMP HISTORY REVEALED NO EVIDENCE OF A 6 UNIT BOLUS BEING PROGRAMMED OR DELIVERED. THE PUMP WAS UNABLE TO BE FULLY INVESTIGATED TO THE UNRESPONSIVE SCREEN.
THE PT REPORTED HEARING A CHIRP, LOOKING AT HIS PUMP AND SEEING THE NORMAL BOLUS SCREEN WITH 6.00 UNITS FLASHING. THE PT REPORTS CANCELLING THE BOLUS AS HE DID NOT WISH TO DELIVER IT. HE WEARS THE PUMP IN A LEATHER CASE CLIPPED TO HIS BELT AND REPORTS NOT BEING ON THE GROUND OR LEANING ON ANYTHING. HE REPORTS NO ISSUES WITH THE BUTTONS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ONE TOUCH PING | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR |