FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1882492 · Received October 21, 2010

Report

Report Number
2531779-2010-01773
Event Type
Malfunction
Date Received
October 21, 2010
Date of Event
September 22, 2010
Report Date
September 22, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. THE DISPLAY SCREEN WAS FOUND TO BE UNRESPONSIVE. THE KEYPAD WAS FOUND TO BE LOOSE AND PEELING. THE PUMP WAS OPENED AND MOISTURE AND CORROSION WERE OBSERVED ON CIRCUIT BOARD ON THE DISPLAY AND THE KEYPAD FLEX CONNECTORS. A REVIEW OF THE PUMP HISTORY REVEALED NO EVIDENCE OF A 6 UNIT BOLUS BEING PROGRAMMED OR DELIVERED. THE PUMP WAS UNABLE TO BE FULLY INVESTIGATED TO THE UNRESPONSIVE SCREEN.

Description of Event or Problem · 1

THE PT REPORTED HEARING A CHIRP, LOOKING AT HIS PUMP AND SEEING THE NORMAL BOLUS SCREEN WITH 6.00 UNITS FLASHING. THE PT REPORTS CANCELLING THE BOLUS AS HE DID NOT WISH TO DELIVER IT. HE WEARS THE PUMP IN A LEATHER CASE CLIPPED TO HIS BELT AND REPORTS NOT BEING ON THE GROUND OR LEANING ON ANYTHING. HE REPORTS NO ISSUES WITH THE BUTTONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1 59 YR