FDA Adverse Event Malfunction Summary report: N

MEDFUSION 3500 SYRINGE INFUSION PUMP

MDR report key: 1882475 · Received October 21, 2010

Report

Report Number
2183502-2010-00469
Event Type
Malfunction
Date Received
October 21, 2010
Date of Event
September 20, 2010
Report Date
October 19, 2010
Manufacturer
SMITHS MEDICAL MD
Product Code
FRN
PMA / PMN Number
K040899
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVAL: THE SUSPECT DEVICE WAS RETURNED AND EVALUATED. TESTING INDICATED THAT THE POSITION POTENTIOMETER WAS INTERMITTENTLY NON-FUNCTIONAL DUE TO A CONTAMINATION BY A FOREIGN MATERIAL. THE POSITION POTENTIOMETER WAS REPLACED. AFTER REPAIR, THE DEVICE WAS FOUND TO PASS ALL DELIVERY, ACCURACY AND FUNCTIONAL TESTS. OUR QUALITY ASSURANCE ENGINEER DETERMINED THAT THE COMPLAINT WAS RELATED TO MAINTENANCE AND USER HANDLING.

Description of Event or Problem · 1

A REPORT WAS RECEIVED THAT STATED THE PUMP ALARMED "CHECK CLUTCH". NO PATIENT INJURY OR ADVERSE EVENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MEDFUSION 3500 SYRINGE INFUSION PUMP SYRINGE INFUSION PUMP FRN SMITHS MEDICAL MD 3500 NA

Patients

Seq Age Sex Outcome Treatment
1 NONE