FDA Adverse Event
Summary report: N
SPACELABS MCARE VITAL SIGNS MONITOR
MDR report key: 1882421
·
Received October 20, 2010
Report
- Report Number
- 3023361-2010-00071
- Date Received
- October 20, 2010
- Date of Event
- October 20, 2010
- Report Date
- October 20, 2010
- Manufacturer
- SPACELABS HEALTHCARE
- Product Code
- MHX
- PMA / PMN Number
- K062095
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TW
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
SPACELABS INVESTIGATION INTO THIS MATTER IS ON-GOING. THE EQUIPMENT IS STILL IN USE AT THE HOSPITAL. NO ONE HAS BEEN INJURED AS A RESULT OF THIS ALLEGED MALFUNCTION. WE WILL PROVIDE A SUPPLEMENTAL REPORT ONCE OUR INVESTIGATION IS COMPLETE.
Description of Event or Problem · 1
SPACELABS RECEIVED A REPORT OF THAT THE NUMERIC HEART RATE VALUE PRINTED ON THE STRIP RECORDING DOES NOT MATCH THE HEART RATE OBTAINED FROM THE WAVEFORM R-R INTERVAL MEASURED ON THE SAME STRIP. THE REPORT INDICATED THAT THE NUMERIC HEART RATE VALUE WAS THE SAME AS THE RESPIRATION RATE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPACELABS MCARE VITAL SIGNS MONITOR | MULTIPARAMETER PATIENT MONITOR | MHX | SPACELABS HEALTHCARE | 91220 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |