FDA Adverse Event Malfunction Summary report: N

COLLEAGUE SINGLE CHANNEL VOLUMETRIC PUMP CE FRENCH

MDR report key: 1882404 · Received October 26, 2010

Report

Report Number
6000001-2010-04490
Event Type
Malfunction
Date Received
October 26, 2010
Date of Event
August 29, 2010
Report Date
September 3, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
Removal / Correction Number
6000001-12/1/08-002-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS DEVICE IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES (US); THEREFORE, IT DOES NOT CONTAIN A US 510K NUMBER. HOWEVER, THIS MDR IS BEING SUBMITTED BECAUSE IT IS THE SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE US. EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH A BATTERY ISSUE WAS CONFIRMED BY BAXTER PERSONNEL AS A BATTERY DEPLETED ALARM. THIS CONDITION WAS CAUSED BY DEPLETED MAIN BATTERIES. THIS CONDITION WAS RESOLVED AT THE FACILITY BY A BAXTER FIELD SERVICE TECHNICIAN BY REPLACING THE MAIN BATTERIES AND BATTERY HARNESS. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). A SERVICE HISTORY REVIEW REVEALED THAT THIS DEVICE WAS PREVIOUSLY SERVICED FOR THE REPORTED CONDITION. DURING PREVIOUS SERVICE, THE PUMPHEAD MODULE, MAIN BATTERY AND SAFETY HARNESS WERE REPLACED. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH MDQ-CAPA-(B)(4).

Description of Event or Problem · 1

IT WAS REPORTED TO BAXTER (B)(4) THAT A COLLEAGUE INFUSION PUMP EXPERIENCED A BATTERY ISSUE. IT IS UNKNOWN WHEN OR IN WHICH CARE AREA THIS EVENT OCCURRED. DURING REVIEW OF THE PUMP'S EVENT HISTORY, BAXTER PERSONNEL CONFIRMED THE REPORTED CONDITION AS A BATTERY DEPLETED ALARM AND DISCOVERED THIS EVENT INTERRUPTED DELIVERY. THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE EVENT IN ASSOCIATION WITH THIS CONDITION. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.08.90.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE SINGLE CHANNEL VOLUMETRIC PUMP CE FRENCH PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1