FDA Adverse Event Malfunction Summary report: N

AUTOMATED PD SET W/CASSETTE4 PRONG

MDR report key: 1882379 · Received October 26, 2010

Report

Report Number
1423500-2010-04934
Event Type
Malfunction
Date Received
October 26, 2010
Date of Event
October 5, 2010
Report Date
October 5, 2010
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE WAS DISCARDED. SHOULD ANY ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THIS COMPLAINT FOR THE SYSTEM ERROR 2240 (AIR IN LINE) THAT OCCURRED DURING DWELL 5 OF 6 WAS NOT CONFIRMED DUE TO A LACK OF SAMPLE. THE CAUSE OF THE COMPLAINT WAS NOT DETERMINED. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Description of Event or Problem · 1

A HOME PATIENT (HP) CONTACTED GLOBAL TECHNICAL SERVICES REGARDING A SYSTEM ERROR (SE) 2240 ALARM THAT OCCURRED ON THE HOME CHOICE (HC) UNIT DURING DWELL 5 OF 6. THE HP WAS STILL CONNECTED, THE SUPPLY BAG WAS EMPTY BUT THERE WAS SOLUTION IN THE HEATER BAG. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) HAD THE HP CYCLE POWER AND EXPLAINED THAT A SE 2240 INDICATES A LARGE AMOUNT OF AIR HAS ENTERED SETUP. THE HP CONFIRMED FINISHING THERAPY WITH MANUALS. THE TSR REVIEWED PROPER PROCEDURES PER THE USER MANUAL WITH THE HP. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION REPORTED. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AUTOMATED PD SET W/CASSETTE4 PRONG SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 69 YR