FDA Adverse Event Injury Summary report: N

HAART 200 AORTIC ANNULOPLASTY DEVICE

MDR report key: 18823317 · Received March 2, 2024

Report

Report Number
3004478276-2024-00119
Event Type
Injury
Date Received
March 2, 2024
Date of Event
December 11, 2023
Report Date
April 18, 2024
Manufacturer
BIOSTABLE
Product Code
KRH
PMA / PMN Number
K171431
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT IS SUBMITTED ON BEHALF OF CORCYM INC. AUTHORIZATION LETTER HAS BEEN SUBMITTED TO FDA THROUGH ESG ACCOUNT CORCYM-LM18. H3 OTHER TEXT : UNKNOWN DISPOSITION.

Additional Manufacturer Narrative · 0

THE MANUFACTURING AND MATERIAL RECORDS FOR THE HAART 200 AORTIC ANNULOPLASTY DEVICE SIZE 25, LOT # 01-01096, INVOLVED IN THE REPORTED EVENT, WERE RETRIEVED AND REVIEWED BY THE MANUFACTURER¿S QUALITY ENGINEERING. THE RESULTS CONFIRMED THAT THIS DEVICE SATISFIED ALL MATERIAL, VISUAL, AND PERFORMANCE STANDARDS REQUIRED FOR A (MODEL # 200-25) HAART 200 AORTIC ANNULOPLASTY DEVICE SIZE 25 AT THE TIME OF MANUFACTURE AND RELEASE. SINCE THE DEVICE INVOLVED IN THE REPORTED EVENT WAS NOT RETURNED FOR ANALYSIS, THE MANUFACTURER COULDN'T PERFORM FURTHER INVESTIGATION ON THE DEVICE. BASED ON THE AVAILABLE INFORMATION, THE DEFINITE ROOT CAUSE OF THE REPORTED EVENT COULD NOT ULTIMATELY BE ESTABLISHED. ACCORDING TO THE CASE DESCRIPTION, IT IS REASONABLE TO ASSUME THAT THE PREDOMINANT FACTOR LEADING TO THE REPORTED EVENT MAY HAVE BEEN A HIGH DEGREE OF STRESS ON THE TISSUE DUE TO THE REPORTED SIGNIFICANT MYOCARDIAL CONTRACTILITY, COMBINED WITH AN UPPER-LIMIT RING SIZE. IN ADDITION, ON THE BASIS OF THE INTRA-OPERATIVE IMAGES CAPTURED AT THE TIME OF THE REOPERATION WHICH WERE PROVIDED TO THE MANUFACTURER, A POTENTIAL INTERFERENCE OF THE DEVICE WITH THE LEAFLETS CANNOT BE RULED OUT. THE MANUFACTURER WILL SUBMIT A FOLLOW UP REPORT IN CASE ANY NEW INFORMATION WILL BE RECEIVED.

Description of Event or Problem · 0

THE MANUFACTURER WAS INFORMED OF AN INTRA-OPERATIVE EXPLANT OF A HAART 200 25 MM DEVICE OCCURRED ON (B)(6) 2023. THE DEVICE HAD BEEN IMPLANTED ON (B)(6) 2023 IN A VERY YOUNG PATIENT WITH BICUSPID NATIVE VALVE AND SEVERE AR ON REGULAR FOLLOW UP SINCE BIRTH. PATIENT WAS A SMOKER AND HAD NO SYMPTOMS. PRE-OPERATIVE TOE PERFORMED IN OCTOBER SHOWED AOR 4+ (FUSED CUSP PROLAPSE, ECCENTRIC JET, 160° - KERCHOVE TYPE A), IN ABSENCE OF STENOSIS AND CALCIFICATION. THE FOLLOWING PARAMETERS WERE ALSO MEASURED: AORTIC ANNULUS 38 MM, ROOT 42 MM AND STJ 32 MM. A PREVIOUS CARDIAC MRI PERFORMED IN (B)(6) 2023 HAD SHOWN THE FOLLOWING VALUES: ANULUS 32 MM, ROOT 41 MM, STJ 31 MM, ASCAO 30 MM, ARCH 25 MM; EF 55%, EDV 258 ML. REPORTEDLY, THE IMPLANT PROCEDURE WAS PERFORMED THROUGH STERNOTOMY WITH NO COMPLICATIONS. THE LIGHTLY CALCIFIED RAPHE WAS SHAVED AND THE NON-FUSED LEAFLET PLICATED WITH TWO STITCHES (PROLENE 6/0). HAART 200 DEVICE WAS IMPLANTED WITH TI-CRON 3/0. THE PATIENT WAS DISCHARGED AT HOME ON THE 5TH POD. THE FOLLOWING PARAMETERS WERE MEASURED THROUGH TTE ON THE 3RD POD: FE 50%, EDV 250 ML. NO AOR, VMAX 1.7 M/SEC. AS REPORTED, AT THE FOLLOW UP PERFORMED 2 MONTHS AFTER THE SURGERY THE FOLLOWING PARAMETERS WERE MEASURED THROUGH TTE/TOE: FE 50%, EDV 256 ML. AOR 4+ (PHT 200 MSEC). A COMPLETE FUSED LEAFLET FLAIL WAS OBSERVED. THE PATIENT WAS CLINICALLY ASYMPTOMATIC. AN AORTIC VALVE REPLACEMENT WITH A MECHANICAL PROSTHESIS (ST. JUDE REGENT 23 MM) WAS THUS PERFORMED ON (B)(6) 2023 THROUGH RE-STERNOTOMY. INTRA-OPERATIVELY, NO HAART DEVICE DEHISCENCE WAS OBSERVED WHILE THE FUSED LEAFLET WAS RUPTURED AT THE BASE, STILL ATTACHED AT THE COMMISSURES. THE SURGERY WAS CONDUCTED WITH NO COMPLICATIONS AND THE PATIENT WAS DISCHARGED AT HOME IN THE 8TH POD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
513063 HAART 200 AORTIC ANNULOPLASTY DEVICE AORTIC ANNULOPLASTY RING KRH BIOSTABLE HAART 200-25 SIZE 25 MM 01-01096
943777 HAART 200 AORTIC ANNULOPLASTY DEVICE AORTIC ANNULOPLASTY RING KRH BIOSTABLE HAART 200-25 SIZE 25 MM 01-01096

Patients

Seq Age Sex Outcome Treatment
1 18 YR Male Required Intervention