FDA Adverse Event Injury Summary report: N

ICL (IMPLANTABLE COLLAMER LENS)

MDR report key: 1882244 · Received October 19, 2010

Report

Report Number
2023826-2010-01058
Event Type
Injury
Date Received
October 19, 2010
Date of Event
August 26, 2010
Report Date
September 20, 2010
Manufacturer
STAAR SURGICAL CO.
Product Code
MTA
PMA / PMN Number
P030016
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4): EVAL CONCLUSION - AT THE CONCLUSION OF THE ORIGINAL INVESTIGATION OPENED FOR THE ISSUE OF ICL VAULTING (BOTH INADEQUATE AND EXCESSIVE), IT WAS DETERMINED THAT THE MOST LIKELY ROOT CAUSE FOR BOTH INADEQUATE AND EXCESSIVE VAULT IS DIFFICULTY IN CLINICAL SIZING. IN NO CASE HAS A RETURNED LENS EXHIBITED A LENGTH MEASUREMENT OUTSIDE THE EXPECTED RANGE ACCORDING TO THE LABELED LENGTH. PRE-OPERATIVE MEASUREMENT TECHNOLOGY AVAILABLE TO CLINICIANS IN THE PAST DID NOT PROVIDE A DIRECT MEASUREMENT OF THE SULCUS, TO WHICH THE ICL SHOULD BE SIZED. THE CURRENT PRACTICE IN SELECTING AN APPROPRIATE ICL FOR A PATIENT IS TO MEASURE WHITE-TO-WHITE AND ANTERIOR CHAMBER DEPTH AND, THROUGH CORRELATIVE ALGORITHMS, PREDICT THE LENGTH OF THE ICL THAT WILL BRIDGE THE SULCUS. THIS METHOD OF SIZING BASED UPON WHITE-TO-WHITE AND ANTERIOR CHAMBER DEPTH MEASUREMENTS WAS USED IN THE FDA CLINICAL TRIAL AND IS RECOMMENDED IN THE CURRENT LABELING OF THE ICL. IF THE PREDICTED LENGTH IS TOO SHORT, THE RESULT MAY BE AN INADEQUATE VAULT. IF THE PREDICTED LENGTH IS TO LONG, THE RESULT MAY BE AN EXCESSIVE VAULT. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SURGEON INSERTED AN ICM120V4 12.0MM IMPLANTABLE COLLAMER LENS ON (B)(6) 2010. THE LENS WAS EXPLANTED ON (B)(6) 2010 DUE TO EXCESSIVE VAULT, ASSOCIATED WITH NARROWING OF THE ANGLE AND SHALLOWING OF THE ACD. ELEVATED IOP WAS NOTED AND AN IRIDECTOMY WAS PERFORMED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ICL (IMPLANTABLE COLLAMER LENS) INTRAOCULAR LENS MTA STAAR SURGICAL CO. ICM120V4 NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention FOAM TIP PLUNGER: MODEL #UNK, LOT #UNK| CARTRIDGE: MODEL #SFC-45 FP, LOT #UNK| INJECTOR: MODEL #MSI-PF UNK, LOT #UNK