FDA Adverse Event Malfunction Summary report: N

TRAPEZOID RX LITHOTRIPTER COMPATIBLE BASKET, MODELS 1086, 1087, 1088

MDR report key: 1882125 · Received October 25, 2010

Report

Report Number
3005099803-2010-04450
Event Type
Malfunction
Date Received
October 25, 2010
Date of Event
August 20, 2010
Report Date
October 1, 2010
Manufacturer
BOSTON SCIENTIFIC - SPENCER
Product Code
LQC
PMA / PMN Number
K040447
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

CONCOMITANT MEDICAL PRODUCT: GUIDEWIRE: (B)(4) VISIGLIDE 0.025 INCH. (B)(4), (SIDE-CAR PUSH-BACK). A VISUAL EXAMINATION OF THE RETURNED DEVICE FOUND RESIDUE PRESENT INDICATING DEVICE USE. FURTHERMORE, IT WAS NOTED THAT THE SIDE-CAR PRESENTED PUSH-BACK AND WAS COLLAPSED. FUNCTIONALLY A GUIDEWIRE WAS LOADED INTO THE SIDE-CAR BUT WAS UNABLE TO PASS. THE CONDITION OF THE RETURNED INCIDENT DEVICE WAS CONSISTENT WITH THE COMPLAINT; GUIDEWIRE COULD NOT BE INSERTED. HOWEVER, DURING DEVICE EVALUATION IT WAS NOTED THAT THE SIDE-CAR PRESENTED PUSH-BACK AND WAS COLLAPSED. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS IT APPEARS THAT THE SIDECAR BECAME COLLAPSED DURING INSERTION INTO THE SCOPE OR ATTEMPTED USE. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) WAS PERFORMED; NO ANOMALIES WERE NOTED. (B)(4).

Description of Event or Problem · 1

IT WAS INITIALLY REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRAPEZOID RX LITHOTRIPTER COMPATIBLE BASKET WAS USED DURING A BILE DUCT STONE REMOVAL PROCEDURE PERFORMED ON (B)(6) 2010. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE GUIDEWIRE COULD NOT BE INSERTED INTO THE RX PORT OF THE DEVICE. THE PROCEDURE WAS COMPLETED WITH ANOTHER TRAPEZOID RX LITHOTRIPTER COMPATIBLE BASKET DEVICE. REPORTEDLY THE DEVICE WAS INSPECTED/TESTED PRIOR TO USE AND NO ANOMALIES WERE NOTED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE GOOD. THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION RESULTS; SIDE-CAR PUSH-BACK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TRAPEZOID RX LITHOTRIPTER COMPATIBLE BASKET, MODELS 1086, 1087, 1088 LITHOTRIPTOR, BILIARY MECHANICAL LQC BOSTON SCIENTIFIC - SPENCER M00510890 13519271

Patients

Seq Age Sex Outcome Treatment
1