HOMECHOICE CYCLER-REFURBISHED
Report
- Report Number
- 1423500-2010-04911
- Event Type
- Malfunction
- Date Received
- October 25, 2010
- Date of Event
- October 6, 2010
- Report Date
- October 6, 2010
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K053512
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4).THE DEVICE HAS BEEN RECEIVED, AND THE EVALUATION IS IN PROCESS. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RECEIVED OPERATIONAL AND IN GOOD CONDITION. THE DEVICE PASSED THE HOMECHOICE RITE (RETURN INSTRUMENT TEST / EVALUATION) FUNCTIONAL TEST BUT FAILED THE RITE ELECTRICAL GROUND BOND TEST. THE PAL (PRODUCT ANALYSIS LAB) EVALUATED THE DEVICE. THERE WERE NO PROBLEMS FOUND DURING AN INTERNAL INSPECTION. THE RESISTANCE BETWEEN THE POWER ENTRY MODULE AND THE DOOR POST GROUND WAS WITHIN THE GROUND BOND SPECIFICATION. THE CAUSE FOR RITE TEST FAILURE - GROUND BOND WAS UNDETERMINED. THE DEVICE'S SERVICE HISTORY RECORD WAS REVIEWED AND NO ISSUES WERE IDENTIFIED THAT MAY HAVE CONTRIBUTED TO THE RITE FAILURE. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS.
DURING EVALUATION OF A RETURNED HOMECHOICE (HC) DEVICE, THE PRODUCT ANALYSIS LABORATORY DETERMINED THE HC MACHINE SYSTEM FAILED RETURNED INSTRUMENT TEST/EVALUATION TESTING DUE THE DEVICE FAILED THE GROUND BOND TEST INDICATING A HIGH RESISTANCE VALUE BETWEEN THE HC AND THE GROUND BOND ON THE POWER SUPPLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECHOICE CYCLER-REFURBISHED | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |