RADIAL JAW 3 BIOPSY FORCEPS
Report
- Report Number
- 3005099803-2010-04459
- Event Type
- Malfunction
- Date Received
- October 25, 2010
- Report Date
- October 1, 2010
- Manufacturer
- BOSTON SCIENTIFIC - COSTA RICA
- Product Code
- FCL
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
A VISUAL EXAMINATION OF THE RETURNED DEVICE FOUND THE NEEDLE BENT. NO ABNORMALITIES WERE NOTED WITH THE DEVICE RIVETING AND WELDING WHICH WERE WITHIN SPECIFICATIONS. FUNCTIONALLY THE DEVICE JAWS WOULD OPEN HOWEVER, THE BENT NEEDLE DID NOT ALLOW PROPER JAW CLOSURE. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) WAS PERFORMED; NO ANOMALIES WERE NOTED. THE CONDITION OF THE RETURNED INCIDENT DEVICE WAS CONSISTENT WITH THE COMPLAINT; NEEDLE BENT. THE MOST PROBABLE ROOT CAUSE IS UNDETERMINABLE DUE TO THE LACK OF EVIDENCE IDENTIFYING THE USE OF THE DEVICE AND SINCE THERE ARE CONTROLS IN THE MANUFACTURING PROCESS WHICH ASSURE PRODUCT INTEGRITY. (B)(4).
THE DEVICE HAS NOT BEEN RECEIVED FOR ANALYSIS. UPON RECEIPT AND COMPLETION OF THE FAILURE ANALYSIS OF THE COMPLAINT DEVICE, IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED.(B)(4)
ACCORDING TO THE COMPLAINANT, FIVE EVENTS OCCURRED DURING SEPARATE PROCEDURES. THE EVENTS WERE EITHER BENT FORCEPS NEEDLE FOUND DURING INSPECTION PRIOR TO USE, OR NO SMOOTH JAW MOVEMENT ENCOUNTERED DURING THE PROCEDURE. THE EVENT "BENT NEEDLE" IS AN MDR REPORTABLE SCENARIO, AND SINCE IT COULD NOT BE ASCERTAINED WHICH EVENT OCCURRED DURING EACH PROCEDURE, FIVE MANUFACTURER REPORTS HAVE BEEN ESTABLISHED. PLEASE REFERENCE REPORT # 3005099803-2010-04458, 3005099803-2010-04459, 3005099803-2010-04460, 3005099803-2010-04461 AND 3005099803-2010-04462. NO VISIBLE DAMAGE WAS REPORTED TO THE PACKAGING OF THE DEVICE AND THE STERILE BARRIER DID NOT APPEAR COMPROMISED. THE PROCEDURE WAS COMPLETED WITH ANOTHER RADIAL JAW 3 SINGLE-USE BIOPSY FORCEPS DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FINE. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION REGARDING THE CIRCUMSTANCES SURROUNDING THIS EVENT HAVE BEEN UNSUCCESSFUL TO DATE. SHOULD ADDITIONAL RELEVANT DETAILS BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
ACCORDING TO THE COMPLAINANT, FIVE EVENTS OCCURRED DURING SEPARATE PROCEDURES. THE EVENTS WERE EITHER BENT FORCEPS NEEDLE FOUND DURING INSPECTION PRIOR TO USE, OR NO SMOOTH JAW MOVEMENT ENCOUNTERED DURING THE PROCEDURE. THE EVENT "BENT NEEDLE" IS AN MDR REPORTABLE SCENARIO, AND SINCE IT COULD NOT BE ASCERTAINED WHICH EVENT OCCURRED DURING EACH PROCEDURE, FIVE MANUFACTURER REPORTS HAVE BEEN ESTABLISHED. PLEASE REFERENCE REPORT # 3005099803-2010-04458, 3005099803-2010-04459, 3005099803-2010-04460, 3005099803-2010-04461 AND 3005099803-2010-04462. NO VISIBLE DAMAGE WAS REPORTED TO THE PACKAGING OF THE DEVICE AND THE STERILE BARRIER DID NOT APPEAR COMPROMISED. THE PROCEDURE WAS COMPLETED WITH ANOTHER RADIAL JAW 3 SINGLE-USE BIOPSY FORCEPS DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FINE. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION REGARDING THE CIRCUMSTANCES SURROUNDING THIS EVENT HAVE BEEN UNSUCCESSFUL TO DATE. SHOULD ADDITIONAL RELEVANT DETAILS BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | RADIAL JAW 3 BIOPSY FORCEPS | FORCEPS, BIOPSY, NON-ELECTRIC | FCL | BOSTON SCIENTIFIC - COSTA RICA | M00515373 | 0013497921 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |