INTEGRATED APD SET W/CASSETTE3-PRONG
Report
- Report Number
- 1423500-2010-04882
- Event Type
- Malfunction
- Date Received
- October 25, 2010
- Date of Event
- September 30, 2010
- Report Date
- September 30, 2010
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
(B)(4). PER THE CUSTOMER THE SAMPLE WAS DISCARDED AND THE LOT NUMBER WAS UNKNOWN, THEREFORE NO EVALUATION OR BATCH REVIEW COULD BE PERFORMED. THIS COMPLAINT IS FOR AIR IN LINE. THIS COMPLAINT WAS NOT CONFIRMED IN THE LAB DUE TO A LACK OF SAMPLE. PER THE COMPLAINT INFORMATION, THERE WAS AIR IN THE PATIENT LINE. THERE WAS NOT ENOUGH DATA WITHIN THE COMPLAINT INFORMATION TO IDENTIFY ROOT CAUSE; THEREFORE THE ROOT CAUSE OF THE COMPLAINT IS NOT DETERMINED. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.
(B)(4). THE LOT NUMBER AND SAMPLE AVAILABILITY ARE UNKNOWN AT THIS TIME. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF BAXTER'S INVESTIGATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.
DURING A FOLLOW UP WITH THE NURSE, IT WAS REVEALED THAT THE ISSUE WAS RESOLVED. THE NURSE ADDED THAT THE HOME PATIENT (HP) WAS SEEN IN THE CLINIC ON (B)(6) 2010 AND IS DOING FINE AND CONTINUING THERAPY. PER NURSE, HP DID NOT HAVE ANY INJURY OR HARM AS A RESULT OF THIS INCIDENT. NO ALLEGATIONS WERE MADE AGAINST ANY OF THE HP'S DIALYSIS PRODUCTS. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS INDICATED. NO FURTHER INFORMATION WAS PROVIDED.
A CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER REPORTING THAT SHE SAW AIR IN THE PATIENT LINE DURING WARMING SOLUTION AND WANTED ASSISTANCE TO END THERAPY ON THE HOMECHOICE (HC). THE TECHNICAL SERVICE REPRESENTATIVE (TSR) ASKED THE CAREGIVER (CG) TO PRESS STOP AND ASSISTED WITH ENDING THERAPY. THE CG WOULD RESTART THE SETUP WITH ALL NEW SUPPLIES. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS INDICATED AT THE TIME OF THE INITIAL REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INTEGRATED APD SET W/CASSETTE3-PRONG | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR |