FDA Adverse Event
Malfunction
Summary report: N
ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM
MDR report key: 1881627
·
Received October 20, 2010
Report
- Report Number
- 2531779-2010-01743
- Event Type
- Malfunction
- Date Received
- October 20, 2010
- Date of Event
- September 20, 2010
- Report Date
- September 20, 2010
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Removal / Correction Number
- 2531779-03/24/2010-003-R
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THERE IS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. A REVIEW OF PUMP HISTORY INDICATED THAT MULTIPLE LOW (NON-ZERO) FORCE LOSS OF PRIME WARNINGS HAD OCCURRED. THE PUMP WAS EXERCISED AND LOST PRIME WITHIN A 24-HOUR PERIOD. DURING EVALUATION THE FORCE SENSOR ASSEMBLY WAS FOUND TO BE OUT OF CALIBRATION. EVALUATION ALSO REVEALED A DISLODGED DISPLAY SCREEN.
Description of Event or Problem · 1
THE PATIENT REPORTED MULTIPLE LOSS OF PRIME WARNINGS FOR THE PAST SEVERAL MONTHS. IT WAS REPORTED THAT THE FREQUENCY INCREASED RECENTLY AND RESULTED IN 5 LOSS OF PRIME WARNINGS IN ONE MORNING. DURING EVALUATION THE FORCE SENSOR WAS FOUND TO BE OUT OF CALIBRATION. EVALUATION ALSO REVEALED A DISLODGED DISPLAY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ONE TOUCH PING | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 41 YR |