FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1881627 · Received October 20, 2010

Report

Report Number
2531779-2010-01743
Event Type
Malfunction
Date Received
October 20, 2010
Date of Event
September 20, 2010
Report Date
September 20, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Removal / Correction Number
2531779-03/24/2010-003-R
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THERE IS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. A REVIEW OF PUMP HISTORY INDICATED THAT MULTIPLE LOW (NON-ZERO) FORCE LOSS OF PRIME WARNINGS HAD OCCURRED. THE PUMP WAS EXERCISED AND LOST PRIME WITHIN A 24-HOUR PERIOD. DURING EVALUATION THE FORCE SENSOR ASSEMBLY WAS FOUND TO BE OUT OF CALIBRATION. EVALUATION ALSO REVEALED A DISLODGED DISPLAY SCREEN.

Description of Event or Problem · 1

THE PATIENT REPORTED MULTIPLE LOSS OF PRIME WARNINGS FOR THE PAST SEVERAL MONTHS. IT WAS REPORTED THAT THE FREQUENCY INCREASED RECENTLY AND RESULTED IN 5 LOSS OF PRIME WARNINGS IN ONE MORNING. DURING EVALUATION THE FORCE SENSOR WAS FOUND TO BE OUT OF CALIBRATION. EVALUATION ALSO REVEALED A DISLODGED DISPLAY.