FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 1881521 · Received October 19, 2010

Report

Report Number
2531779-2010-01745
Event Type
Malfunction
Date Received
October 19, 2010
Date of Event
September 20, 2010
Report Date
September 20, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS FOR EVAL. EVAL REVEALED THAT THE PUMP WAS INTERMITTENTLY UNRESPONSIVE TO UP AND DOWN ARROW BUTTON PRESSES. THE UP AND DOWN ARROW BUTTON CONTACTS WERE OBSERVED TO BE MISALIGNED.

Description of Event or Problem · 1

THE REPORTER CONTACTED ANIMAS ON BEHALF OF THE PT (HER DAUGHTER) ALLEGED THAT THE PUMP WAS UNRESPONSIVE TO KEYPAD BUTTON PRESSES. THE REPORTER CLAIMED THAT SHE HAD TO PRESS THE UP AND DOWN ARROW BUTTONS HARD AND SEVERAL TIMES IN ORDER FOR THE PUMP TO RESPOND. THE REPORTER ALSO CLAIMED THAT THE PUMP WAS RESPONDING INTERMITTENTLY TO OK BUTTON PRESSES. SHE ALSO ALLEGED THAT THE BUTTONS WOULD NOT CLICK OR SPRING BACK AS USUAL. THE REPORTER DENIED THAT THE DEVICE WAS EXPOSED TO WATER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORP. ANIMAS 2020 NA

Patients

Seq Age Sex Outcome Treatment
1 14 YR