FDA Adverse Event
Summary report: N
WARMING LAMP
MDR report key: 1881518
·
Received October 18, 2010
Report
- Report Number
- 1881518
- Date Received
- October 18, 2010
- Date of Event
- October 8, 2010
- Report Date
- October 11, 2010
- Manufacturer
- PHILIPS RESPIRONICS (FORMERLY J.H. EMERSON CO.)
- Product Code
- ILY
- Report Source
- User Facility report
- Reporter Location
- DC, US
Narratives
Description of Event or Problem · 1
A PEDIATRIC PATIENT (IN PATIENT) RECEIVED A 2ND DEGREE BURN FROM A HEAT LAMP. ====================== HEALTH PROFESSIONAL'S IMPRESSION======================THE DEVICE WAS PLACED TOO CLOSE TO THE PATIENT (MORE THAN 28 CM PER MANUFACTURER RECOMMENDATIONS)
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | WARMING LAMP | LAMP, INFRARED, THERAPEUTIC HEATING | ILY | PHILIPS RESPIRONICS (FORMERLY J.H. EMERSON CO.) | 96-TDL | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * |