FDA Adverse Event
Malfunction
Summary report: N
ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM
MDR report key: 1881512
·
Received October 19, 2010
Report
- Report Number
- 2531779-2010-01754
- Event Type
- Malfunction
- Date Received
- October 19, 2010
- Date of Event
- September 20, 2010
- Report Date
- September 20, 2010
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UT, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE PUMP HAS BEEN RETURNED TO ANIMAS FOR EVAL. EVAL REVEALED THAT THE PUMP WAS UNRESPONSIVE TO DOWN ARROW BUTTON PRESSES. ALL OTHER KEYPAD BUTTONS WERE RESPONSIVE TO USER INPUTS. EVIDENCE OF MOISTURE CORROSION WAS OBSERVED UNDER ALL KEY CONTACTS. THE KEYPAD WAS FOUND TO BE TORN AND LIFTED BY THE OK BUTTON.
Description of Event or Problem · 1
THE PT CONTACTED ANIMAS ALLEGING THAT THE PUMP WAS NOT RESPONDING TO DOWN ARROW BUTTON PRESSES. THE REPORTER CLAIMED THAT THE BUTTON WAS NO LONGER SPRINGING BACK, BUT WAS CLICKING AS USUAL. THE KEYPAD WAS ALSO REPORTEDLY PEELING BY THE OK BUTTON. THE PT DENIED THAT THE DEVICE WAS EXPOSED TO MOISTURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ONE TOUCH PING | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 23 YR |