FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1881512 · Received October 19, 2010

Report

Report Number
2531779-2010-01754
Event Type
Malfunction
Date Received
October 19, 2010
Date of Event
September 20, 2010
Report Date
September 20, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS FOR EVAL. EVAL REVEALED THAT THE PUMP WAS UNRESPONSIVE TO DOWN ARROW BUTTON PRESSES. ALL OTHER KEYPAD BUTTONS WERE RESPONSIVE TO USER INPUTS. EVIDENCE OF MOISTURE CORROSION WAS OBSERVED UNDER ALL KEY CONTACTS. THE KEYPAD WAS FOUND TO BE TORN AND LIFTED BY THE OK BUTTON.

Description of Event or Problem · 1

THE PT CONTACTED ANIMAS ALLEGING THAT THE PUMP WAS NOT RESPONDING TO DOWN ARROW BUTTON PRESSES. THE REPORTER CLAIMED THAT THE BUTTON WAS NO LONGER SPRINGING BACK, BUT WAS CLICKING AS USUAL. THE KEYPAD WAS ALSO REPORTEDLY PEELING BY THE OK BUTTON. THE PT DENIED THAT THE DEVICE WAS EXPOSED TO MOISTURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1 23 YR