FDA Adverse Event
Malfunction
Summary report: N
TERUMO CDI 500 BLOOD PARAMETER MONITOR
MDR report key: 1881473
·
Received October 8, 2010
Report
- Report Number
- 1828100-2010-01696
- Event Type
- Malfunction
- Date Received
- October 8, 2010
- Date of Event
- September 15, 2010
- Report Date
- October 8, 2010
- Manufacturer
- TERUMO CARDIOVASCULAR SYSTEMS CORP.
- Product Code
- DRY
- PMA / PMN Number
- K972962
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Description of Event or Problem · 1
DURING ROUTINE TESTING OF THE DEVICE AT THE SERVICE CENTER, THE USER OBSERVED AN INTERMITTENT "COLOR CHIP FAILURE" ERROR MESSAGE. THE COLOR CHIP FAILURE WAS FOUND TO BE CAUSED BY A BROKEN CLIP ON THE HEMOCRIT SATURATION PROBE. THE MONITOR WAS ORIGINALLY RETURNED FOR REPAIR DUE TO A FAILURE TO BOOT UP WHEN THE COLOR CHIP FAILURE WAS FOUND. SINCE THE EVENT OCCURRED DURING ROUTINE TESTING OF THE DEVICE, THERE WAS NO PT INVOLVEMENT DURING THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TERUMO CDI 500 BLOOD PARAMETER MONITOR | ON LINE BLOOD GAS MONITOR | DRY | TERUMO CARDIOVASCULAR SYSTEMS CORP. | 500AHCT |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |