FDA Adverse Event Injury Summary report: N

MECTACER BIOLOX DELTA CERAMIC BALL HEAD 12/14 Ø 36 SIZE S - 4

MDR report key: 18814606 · Received March 1, 2024

Report

Report Number
3005180920-2024-00067
Event Type
Injury
Date Received
March 1, 2024
Date of Event
January 31, 2024
Report Date
March 1, 2024
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
UDI-DI
07630030809194
PMA / PMN Number
K112115
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

BATCH REVIEW PERFORMED ON 05-FEB-2024: LOT 2315550: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 30-JUN-2023. EXPIRATION DATE: 2028-06-10. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED CASE DURING THE PERIOD OF THE REVIEW. LOT 2309707: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 03-JUL-2023. EXPIRATION DATE: 2028-06-05. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED CASE DURING THE PERIOD OF REVIEW.

Description of Event or Problem · 0

AT ABOUT 3 MONTHS AFTER PRIMARY, THE PATIENT CAME IN REPORTING PAIN DUE TO A DISLOCATION OF THE HEAD FROM THE LINER. THE SURGEON REVISED THE HEAD AND LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
307172 MECTACER BIOLOX DELTA CERAMIC BALL HEAD 12/14 Ø 36 SIZE S - 4 HIP CERAMIC HEAD LZO MEDACTA INTERNATIONAL SA 2315550 07630030809194

Patients

Seq Age Sex Outcome Treatment
1 48 YR Female Required Intervention