FDA Adverse Event
Malfunction
Summary report: N
840 VENTILATOR
MDR report key: 1881254
·
Received October 6, 2010
Report
- Report Number
- 8020893-2010-00491
- Event Type
- Malfunction
- Date Received
- October 6, 2010
- Date of Event
- September 6, 2010
- Report Date
- September 16, 2010
- Manufacturer
- PURITAN BENNETT CORP.
- Product Code
- CBK
- PMA / PMN Number
- K970460
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Description of Event or Problem · 1
THE SERVICE REPORT SHOWS THE CUSTOMER REPORTED THAT THE 840 VENTILATOR STOPPED CYCLING WHILE IN USE ON A PT. THE PT WAS NOT HARMED OR INJURED AS A RESULT OF THE EVENT. THE NELLCOR PURITAN BENNETT CUSTOMER SUPPORT ENGINEER (CSE) INSPECTED THE DEVICE AND COULD NOT DUPLICATE THE ALLEGED EVENT. THE UNIT PASSED EXTENDED SELF-TESTING AND NO PARTS WERE REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 840 VENTILATOR | VENTILATOR | CBK | PURITAN BENNETT CORP. | 840 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |