FDA Adverse Event Malfunction Summary report: N

QUADROX D

MDR report key: 1881248 · Received October 6, 2010

Report

Report Number
8010762-2010-14610
Event Type
Malfunction
Date Received
October 6, 2010
Date of Event
July 29, 2010
Report Date
August 2, 2010
Manufacturer
MAQUET CARDIOPULMONARY AG
Product Code
DTZ
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

MAQUET CARDIOVASCULAR SUBMITS THIS REPORT ON BEHALF OF THE LEGAL MANUFACTURER OF THE DEVICE MAQUET CARDIOPULMONARY AG, (B)(4). WE CONFIRMED THAT THERE ARE CRACKS AT THE VENOUS AND THE ARTERIAL SIDES OF THE HOUSING. ACCORDING TO THE CUSTOMER'S INFO A ROLLER PUMP WAS USED AND THE OXYGENATOR WORKED FINE UP TO SIX OR SEVEN DAYS, HOWEVER, OUR OXYGENATORS HMOD 2030 AND BBQ-HMOD HAVE AN INDICATIONS FOR USE FOR ONLY UP TO SIX HOURS. WE HAVE TRIED TO SIMULATE THIS FAILURE IN THE LABORATORY WITH LONG TERM TESTS USING THE ROLLER PUMP. THE TESTS OF 4 OXYGENATORS RAN WITH ROLLER PUMPS DURING 20 DAYS AT A TEMPERATURE OF 37 DEGREE C, A PRESSURE OF 500 MMHG AND A WATER FLOW OF 4 1/MIN. NO CRACKS COULD BE OBSERVED AFTER AND DURING RUNNING THE TESTS. THEREFORE, THE ROOT CAUSE COULD NOT BE CONFIRMED. HOWEVER, WE ARE CONFIDENT THAT THE MOST PROBABLE ROOT CAUSE FOR THESE CRACKS IS THE PROLONGED USE IN COMBINATION WITH USING A ROLLER PUMP, WHERE THE OXYGENATOR MAY BECOME STRESSED DUE TO PULSATION OVER TIME. OUR BASIS FOR THIS IS THAT MAQUET CARDIOPULMONARY AG HAS CE CERTIFIED SYSTEMS FOR LONG-TERM USE OUTSIDE THE U.S. WHICH ONLY INCLUDE CENTRIFUGAL PUMPS RATHER THAN ROLLER PUMPS AND PROBLEMS WITH CRACKS ARE NOT KNOWN WITH THESE SPECIAL TUBING SETS. WE BELIEVE THAT THIS TYPE OF FAILURE CAN BE AVOIDED IN THE FUTURE BY THE CUSTOMER'S ADHERENCE TO THE PRODUCT'S INSTRUCTIONS FOR USE WHICH SPECIFIES THAT THE UTILIZATION PERIOD IS RESTRICTED TO SIX HOURS. (B)(4).

Description of Event or Problem · 1

OXYGENATOR DEVELOPED A CRACK ON THE HOUSING AFTER SIX OR SEVEN DAYS OF USE. THE EQUIPMENT WAS CHANGED OUT. THERE WERE NO LEAKS NOTICED AND THERE WAS NO ADVERSE EFFECT TO THE PT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 QUADROX D OXYGENATOR SYSTEM DTZ MAQUET CARDIOPULMONARY AG HMOD 2030 70041436

Patients

Seq Age Sex Outcome Treatment
1 15 YR Other