FDA Adverse Event Injury Summary report: N

CIAGLIA BLUE DOLPHIN PERCUTANEOUS TRACHEOSTOMY INTRODUCER SET

MDR report key: 1881086 · Received October 20, 2010

Report

Report Number
1820334-2010-00511
Event Type
Injury
Date Received
October 20, 2010
Date of Event
September 20, 2010
Report Date
September 20, 2010
Manufacturer
COOK INC.
Product Code
JOH
PMA / PMN Number
K072148
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVENT IS CURRENTLY UNDER INVESTIGATION.

Description of Event or Problem · 1

WHILE THE PHYSICIAN TRIED TO INFLATE THE BALLOON A SECOND TIME (AFTER ONE FAILED INSERTION OF THE TRACHEOSTOMY TUBE), THE BALLOON BURST AT A PRESSURE OF 11 ATM, WHILE STILL INSIDE THE PATIENT. THE PHYSICIAN LEFT THE WIRE GUIDE INSIDE THE PATIENT, USED A NEW BLUE RHINO SET TO INSERT A TRACHEOSTOMY CANNULA AND PLACED IT. THE PHYSICIAN THEN REMOVED THE SMALL PIECES FROM THE BALLOON VIA A BRONCHOSCOPY FROM THE TRACHEA OF THE PATIENT. PATIENT OUTCOME WAS NOT PROVIDED BY REPORTER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CIAGLIA BLUE DOLPHIN PERCUTANEOUS TRACHEOSTOMY INTRODUCER SET JOH TUBE TRACHEOSTOMY AND TUBE CUFF JOH COOK INC. NA F2462299

Patients

Seq Age Sex Outcome Treatment
1 60 YR Required Intervention