FDA Adverse Event Death Summary report: N

ULTRAFLEX ESOPHAGEAL NG STENT SYSTEM

MDR report key: 1880975 · Received October 25, 2010

Report

Report Number
3005099803-2010-04522
Event Type
Death
Date Received
October 25, 2010
Date of Event
October 4, 2010
Report Date
October 4, 2010
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
ESW
PMA / PMN Number
K091816
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4) (NO CODE AVAILABLE) FOR THE REPORTED ISSUE OF STENT PARTIALLY DEPLOYED.THE DEVICE HAS BEEN RECEIVED FOR ANALYSIS. UPON COMPLETION OF THE FAILURE ANALYSIS OF THE COMPLAINT DEVICE, IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED.

Additional Manufacturer Narrative · 1

A VISUAL EXAMINATION OF THE RETURNED DEVICE FOUND THAT THE PROXIMAL END OF THE STENT WAS PARTIALLY DEPLOYED SO THAT APPROXIMATELY 25MM OF THE DISTAL END OF THE STENT WAS UNDEPLOYED. DURING ANALYSIS THE REMAINDER OF THE STENT WAS FULLY DEPLOYED WITHOUT ANY RESTRICTION OR ISSUES BEING NOTED. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) WAS PERFORMED; NO ANOMALIES WERE NOTED THAT COULD BE RELATED TO THIS COMPLAINT. A REVIEW OF COMPLAINT HISTORY FOR THE REPORTED LOT NUMBER WAS PERFORMED AND CONCLUDED THERE WERE NO OTHER COMPLAINTS REPORTED FOR THIS LOT. A LABELING REVIEW WAS PERFORMED, AND FROM THE INFORMATION AVAILABLE THIS DEVICE WAS NOT USED PER THE DIRECTIONS FOR USE (DFU) / PRODUCT LABEL. THE ULTRAFLEX COVERED AND UNCOVERED ESOPHAGEAL AND ESOPHAGEAL NG STENT SYSTEM IS INTENDED FOR MAINTAINING ESOPHAGEAL LUMINAL PATENCY IN ESOPHAGEAL STRICTURES CAUSED BY INTRINSIC AND/OR EXTRINSIC MALIGNANT TUMORS ONLY. THE ULTRAFLEX COVERED ESOPHAGEAL AND THE ESOPHAGEAL NG STENT SYSTEM ARE ALSO BOTH INDICATED FOR OCCLUSION OF CONCURRENT ESOPHAGEAL FISTULA. THIS STENT WAS PLACED TO TREAT AN IATROGENIC LEAK AND NOT TO TREAT A STRICTURE. THE MOST PROBABLE ROOT CAUSE IS USE/USER ERROR.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT TWO ULTRAFLEX ESOPHAGEAL COVERED STENTS WERE USED DURING A PROCEDURE WITHIN A PATIENT WITH A STRICTURE ON (B)(6) 2010. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE FIRST STENT (THE SUBJECT OF MANUFACTURER REPORT # 3005099803-2010-04522) WAS POSITIONED ABOVE THE PYLORUS IN THE ESOPHAGUS. DURING DEPLOYMENT, AFTER SEVENTY FIVE PERCENT OF THE STENT WAS DEPLOYED, RESISTANCE WAS ENCOUNTERED AND THE STENT WAS UNABLE TO BE IMPLANTED. THE PARTIALLY DEPLOYED STENT WAS REMOVED FROM THE PATIENT. A SECOND ULTRAFLEX ESOPHAGEAL COVERED STENT (THE SUBJECT OF MANUFACTURER REPORT # 3005099803-2010-04523) WAS IMPLANTED SUCCESSFULLY. TWENTY FOUR HOURS POST PROCEDURE, THE STENT WAS VISUALLY EXAMINED, AND THE PHYSICIAN BELIEVED THAT THE STENT COVERING WAS "BROKEN". ADDITIONAL INFORMATION WAS REPORTED TO BOSTON SCIENTIFIC, THAT THE PATIENT HAS SINCE PASSED AWAY. DESPITE NUMEROUS ATTEMPTS BOSTON SCIENTIFIC HAS BEEN UNABLE OBTAIN ADDITIONAL INFORMATION REGARDING THE CIRCUMSTANCES SURROUNDING THIS EVENT. SHOULD ADDITIONAL RELEVANT DETAILS BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT TWO ULTRAFLEX ESOPHAGEAL COVERED STENTS WERE USED DURING A PROCEDURE WITHIN A PATIENT WITH A STRICTURE ON (B)(6) 2010. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE THE FIRST STENT (THE SUBJECT OF MANUFACTURER REPORT # 3005099803-2010-04522) WAS POSITIONED ABOVE THE PYLORUS IN THE ESOPHAGUS. DURING DEPLOYMENT, AFTER SEVENTY FIVE PERCENT OF THE STENT WAS DEPLOYED, RESISTANCE WAS ENCOUNTERED AND THE STENT WAS UNABLE TO BE IMPLANTED. THE PARTIALLY DEPLOYED STENT WAS REMOVED FROM THE PATIENT. A SECOND ULTRAFLEX ESOPHAGEAL COVERED STENT (THE SUBJECT OF MANUFACTURER REPORT # 3005099803-2010-04523) WAS IMPLANTED SUCCESSFULLY. TWENTY FOUR HOURS POST PROCEDURE, THE STENT WAS VISUALLY EXAMINED, AND THE PHYSICIAN BELIEVED THAT THE STENT COVERING WAS "BROKEN." ADDITIONAL INFORMATION WAS REPORTED TO BOSTON SCIENTIFIC, THAT THE PATIENT HAS SINCE PASSED AWAY. DESPITE NUMEROUS ATTEMPTS BOSTON SCIENTIFIC HAS BEEN UNABLE OBTAIN ADDITIONAL INFORMATION REGARDING THE CIRCUMSTANCES SURROUNDING THIS EVENT. SHOULD ADDITIONAL RELEVANT DETAILS BECOME AVAILABLE A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ***THE FOLLOWING INFORMATION WAS REPORTED TO BOSTON SCIENTIFIC ON (B)(4) 2010.*** DURING THE ESOPHAGOGASTRODUODENOSCOPY PROCEDURE ON (B)(6) 2010 THE PATIENT'S ANATOMY WAS NOT DILATED. THERE WERE NO ISSUES REMOVING THE PARTIALLY DEPLOYED STENT FROM THE PATIENT, (THE SUBJECT OF MANUFACTURER REPORT # 3005099803-2010-04522). THE INDICATION FOR THE STENT PLACEMENT WAS NOT A STRICTURE, BUT AN IATROGENIC GASTRIC LEAK. AFTER THE STENT WAS PLACED (THE SUBJECT OF MANUFACTURER REPORT # 3005099803-2010-04523), LEAKAGE STOPPED FASTER; HOWEVER LEAKAGE REOCCURRED 24 HOURS LATER. ON (B)(6) 2010, 48 HOURS AFTER STENT PLACEMENT, AN ENDOSCOPIC PROCEDURE SHOWED EXTENSIVE DAMAGE/DISSOLUTION TO THE STENT COVERING. A SECOND ULTRAFLEX STENT WAS PLACED TO RESOLVE THE EVENT. THERE WAS NO MALFUNCTION OF THE SECOND ULTRAFLEX STENT. AFTER 4 DAYS, FURTHER TREATMENT WAS STOPPED DUE TO "INCURABLE MALIGNANCY" AND THE 'DETERIORATING" CONDITION OF THE PATIENT. ON (B)(6) 2010 THE PATIENT PASSED AWAY. NO AUTOPSY WAS PERFORMED; HOWEVER THE CAUSE OF DEATH WAS REPORTED AS "GENERAL MALAISE." IN THE PHYSICIAN'S ASSESSMENT, LEAKAGE SINCE (B)(6), MAY HAVE CONTRIBUTED TO THE DECREASE OF THE PATIENT'S GENERAL CONDITION. IN THE PHYSICIAN'S ASSESSMENT, THE DAMAGE TO THE COVERING OF THE STENT DID CONTRIBUTE TO THE PATIENT'S DEATH. IT WAS REPORTED THAT THE PATIENT HAD MULTIPLE CONTRIBUTING CO-MORBIDITIES. IT IS IMPORTANT TO NOTE THAT THE ULTRAFLEX COVERED AND UNCOVERED ESOPHAGEAL AND ESOPHAGEAL NG STENT SYSTEM IS INTENDED FOR MAINTAINING ESOPHAGEAL LUMINAL PATENCY IN ESOPHAGEAL STRICTURES CAUSED BY INTRINSIC AND/OR EXTRINSIC MALIGNANT TUMORS ONLY. THE ULTRAFLEX COVERED ESOPHAGEAL AND THE ESOPHAGEAL NG STENT SYSTEM ARE ALSO BOTH INDICATED FOR OCCLUSION OF CONCURRENT ESOPHAGEAL FISTULA.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ULTRAFLEX ESOPHAGEAL NG STENT SYSTEM PROSTHESIS, ESOPHAGEAL ESW BOSTON SCIENTIFIC - GALWAY M00514250 13121806

Patients

Seq Age Sex Outcome Treatment
1 Death