FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES

MDR report key: 18809629 · Received February 29, 2024

Report

Report Number
1024879-2024-00182
Event Type
Malfunction
Date Received
February 29, 2024
Date of Event
January 29, 2024
Report Date
May 8, 2024
Manufacturer
BECTON, DICKINSON & CO., (BD)
Product Code
JKA
UDI-DI
50382903679831
PMA / PMN Number
BK050036
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.6 INVESTIGATION SUMMARY: MATERIAL #:367983, LOT/BATCH #: 3153025. BD HAD NOT RECEIVED SAMPLES, BUT ONE (1) PHOTO WAS PROVIDED FOR INVESTIGATION. THE PHOTO WAS REVIEWED AND THE INDICATED FAILURE MODE FOR UNDERFILL WAS OBSERVED. ADDITIONALLY, TWENTY (20) RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY FUNCTIONAL TESTING TO DRAW TEST FOR LOW OR NO DRAW AND THE ISSUE OF UNDERFILL WAS OBSERVED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE INDICATED FAILURE MODE OF UNDERFILLED. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. H3 OTHER TEXT : SEE H.10.

Additional Manufacturer Narrative · 0

(B)(6). H3. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING COLLECTION WITH A BD VACUTAINER® SST¿ BLOOD COLLECTION TUBE, THERE WAS INSUFFICIENT NEGATIVE PRESSURE. THERE WERE NO REPORTED ADVERSE EFFECTS ON PATIENTS.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING COLLECTION WITH A BD VACUTAINER® SST¿ BLOOD COLLECTION TUBE, THERE WAS INSUFFICIENT NEGATIVE PRESSURE. THERE WERE NO REPORTED ADVERSE EFFECTS ON PATIENTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
497328 BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON & CO., (BD) 3153025 50382903679831

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown