FDA Adverse Event Malfunction Summary report: N

THERAPY ABLATION CATHETER

MDR report key: 1880876 · Received October 1, 2010

Report

Report Number
2030404-2010-00167
Event Type
Malfunction
Date Received
October 1, 2010
Date of Event
September 2, 2010
Report Date
September 3, 2010
Manufacturer
IRVINE BIOMEDICAL, A ST. JUDE MEDICAL COMPANY
Product Code
OAD
PMA / PMN Number
P060019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATIONS HAVE CONFIRMED A LEAK IN THE HANDLE OF THE CATHETER IS CAUSED BY THE LUMEN BREAKING AT THE EDGE OF THE CATHETER IN THE PROTECTING TUBE. REVIEW OF THE DEVICE HISTORY RECORDS CONFIRMED THIS LOT MET MFG REQUIREMENTS PRIOR TO SHIPMENT. A QUALITY IMPROVEMENT PROJECT WAS INITIATED TO ADDRESS THIS ISSUE. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT SALINE WAS LEAKING AT THE HANDLE.