FDA Adverse Event Malfunction Summary report: N

ENDOSTAT ELECTROSURGICAL UNIT

MDR report key: 1880630 · Received October 25, 2010

Report

Report Number
3005099803-2010-04420
Event Type
Malfunction
Date Received
October 25, 2010
Date of Event
September 22, 2010
Report Date
September 29, 2010
Manufacturer
BOSTON SCIENTIFIC - MARLBOROUGH
Product Code
KNS
PMA / PMN Number
K913881
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE RETURNED DEVICE PRESENTED WITH SOME MINOR SCRATCHES ON THE COVER, BUT WAS OTHERWISE IN FAIR PHYSICAL CONDITION. ALL KNOBS AND SWITCHES ON THE UNIT FUNCTIONED AS INTENDED, AND ELECTRICALLY, THE UNIT WAS WITHIN SPECIFICATION. ALL CONNECTIONS WITHIN THE UNIT WERE CHECKED AND VERIFIED TO BE IN GOOD WORKING ORDER. THE TUBING OF THE PUMP WAS CLAMPED OFF WHILE THE IRRIGATION WAS ACTIVATED, AND THE PUMP STILL FUNCTIONED PROPERLY. THE FOOTSWITCH WIRE WAS ALSO MANIPULATED TO CHECK FOR POTENTIAL ISSUES, BUT NONE WERE FOUND. THE CONDITION OF THE RETURNED DEVICE WAS NOT CONSISTENT WITH THE COMPLAINT OF PUMPING ISSUES. THE COMPLAINT WAS NOT CONFIRMED. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) WAS PERFORMED; NO ANOMALIES WERE NOTED.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOSTAT II ELECTROSURGICAL UNIT WAS USED DURING A COLONOSCOPY PROCEDURE ON (B)(6), 2010. ACCORDING TO THE COMPLAINANT, THE IRRIGATION SYSTEM ON THE UNIT DID NOT PROPERLY CIRCULATE FLUID; THE PUMP TURNED VERY SLOWLY. IN ORDER TO CIRCUMVENT THE PROBLEM, THE COMPLAINANT USED A 60CC SYRINGE TO INJECT FLUID. THERE WERE NO REPORTED PATIENT ISSUES AS A RESULT OF THIS EVENT. THE PROCEDURE WAS COMPLETE WITH THIS DEVICE. AT THE CONCLUSION OF THE PROCEDURE, THE COMPLAINANT NOTED THAT A RUBBER MECHANISM WITHIN THE PUMP APPEARED TO BE WORN DOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOSTAT ELECTROSURGICAL UNIT UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) KNS BOSTON SCIENTIFIC - MARLBOROUGH M00540800

Patients

Seq Age Sex Outcome Treatment
1