FDA Adverse Event Malfunction Summary report: N

CE INFUSOR LV10, 12 PACK

MDR report key: 1880564 · Received October 22, 2010

Report

Report Number
6000001-2010-04439
Event Type
Malfunction
Date Received
October 22, 2010
Date of Event
October 1, 2010
Report Date
October 1, 2010
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
MEB
PMA / PMN Number
K041738
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NZ
Reporter Occupation
PHARMACIST

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE CUSTOMER DISCARDED THE DEVICE. THEREFORE, THE REPORTED CONDITION OF A RUPTURED RESERVOIR COULD NOT BE CONFIRMED. THE ROOT CAUSE FOR RUPTURED RESERVOIRS IS CURRENTLY BEING INVESTIGATED UNDER CAPA# IDC-CAPA-(B)(4). A BATCH REVIEW WAS PERFORMED FINDING THAT NO EXCEPTION/NON-CONFORMANCE REPORTS WERE DOCUMENTED FOR THIS LOT. ALL RELEASE CRITERIA WERE MET FOR THE BUILD OF THE LOT. SHOULD THE SAMPLE AND/OR ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED TO BAXTER (B)(4) THAT THE RESERVOIR OF AN INFUSOR LV 10 DEVICE RUPTURED BEFORE USE. THE DEVICE WAS FILLED WITH A 240ML SOLUTION OF BENZYLPENICILLIN, AT 30MG PER ML, AND NORMAL SALINE AT THE TIME OF THE RUPTURE. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION NECESSARY AS THIS EVENT DID NOT OCCUR DURING PATIENT USE. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CE INFUSOR LV10, 12 PACK PUMP, INFUSION, ELASTOMERIC MEB BAXTER HEALTHCARE - IRVINE 10H018