FDA Adverse Event Injury Summary report: N

VASOVIEW HEMOPRO EVH SYSTEM

MDR report key: 1880341 · Received October 20, 2010

Report

Report Number
2242352-2010-03132
Event Type
Injury
Date Received
October 20, 2010
Report Date
September 21, 2010
Manufacturer
MAQUET CARDIOVASCULAR, LLC.
Product Code
GEI
PMA / PMN Number
K052274
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO MAQUET CARDIAC SURGERY ON (B)(6) 2010, FOR INVESTIGATION. VISUAL INSPECTION REVEALED THAT THERE WERE NO NONCONFORMITIES WITH THE DEVICE. THERE WAS NO EVIDENCE OF BLOOD. A PRE-CAUTERY TEST WAS PERFORMED ON THE DEVICE WITH A REFERENCE POWER SUPPLY AND EXTENSION CABLE. THE DEVICE PASSED THE PRE-CAUTERY TEST; IT PRODUCED ENERGY AND STEAM, AND DID NOT REMAIN ACTIVATED. BASED UPON THESE FINDINGS, THE REPORTED FAILURE, "CUTTING BUT NOT SEALING" COULD NOT BE CONFIRMED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE PRODUCT. THERE WAS NO NONCONFORMANCE WHICH COULD HAVE CONTRIBUTED TO THIS FAILURE MODE. (B)(4).

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE ABOUT THREE MONTHS PRIOR TO THE DATE OF THIS REPORT, THE VASOVIEW HEMOPRO DEVICE WAS CUTTING BUT NOT SEALING, WHICH LED TO SOME BLEEDING. A NEW KIT WAS USED TO STOP THE BLEEDING WITH NO OTHER PT EFFECTS REPORTED. THE REPORTER STATED THIS EVENT CAME UP IN CONVERSATION WHEN DISCUSSING A DIFFERENT EVENT. A NEW KIT WAS OPENED TO COMPLETE THE PROCEDURE. THE PRODUCT WAS RETURNED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VASOVIEW HEMOPRO EVH SYSTEM ENDOSCOPIC VESSEL HARVESTING SYSTEM GEI MAQUET CARDIOVASCULAR, LLC. VH-3000 25010352

Patients

Seq Age Sex Outcome Treatment
1 NA Other