FDA Adverse Event Injury Summary report: N

ASAHI ASTATO 30 PERIPHERAL GUIDEWIRE

MDR report key: 1880316 · Received October 20, 2010

Report

Report Number
3003775027-2010-00025
Event Type
Injury
Date Received
October 20, 2010
Date of Event
September 22, 2010
Report Date
September 25, 2010
Manufacturer
ASAHI INTECC CO., LTD.
Product Code
DQX
PMA / PMN Number
K071721
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE GUIDEWIRE WAS BROKEN OFF AT APPROX 70 CM FROM TIP END. IT IS PRESUMED THAT THE BENDING FORCE WAS REPEATEDLY APPLIED TO THE GUIDEWIRE SHAFT THAT WAS LOCATED AND BENT IN THE BIFURCATION OF COMMON ILIAC ARTERY, WHEN PHYSICIAN WAS REPEATEDLY PUSHING-PULLING THE GUIDEWIRE IN ORDER TO TRY TO CROSS THE HEAVILY CALCIFIED LESION IN TP-TRUNK, AND THAT THE ACCUMULATED FORCE EXCEEDED THE GUIDEWIRE DESIGN LIMIT WHEN THE SHAFT BREAKAGE OCCURRED.

Description of Event or Problem · 1

TO A HEAVILY CALCIFIED CTO IN THE TIBIAL-PERONEAL TRUNK IN RIGHT LEG, PUNCTURE WAS MADE IN THE LEFT GROIN AND APPROACH WAS MADE UP-AND-OVER THE COMMON ILIAC ARTERY. PHYSICIAN WAS ATTEMPTING TO CROSS A HEAVILY CALCIFIED LESION, AND WAS ON HIS 3RD ASTATO GUIDE WIRE BECAUSE THE PREVIOUS WIRE TIPS HAD BECOME DEFORMED DUE TO MULTIPLE ATTEMPTS TO CROSS CTO. THIRD WIRE ADVANCED THROUGH THE OTHER COMPANY'S SUPPORT CATHETER AND A GUIDING SHEATH. AS HE MADE CONTACT WITH CTO TO CROSS, THE PROXIMAL PORTION OF THE WIRE BROKE OFF IN THE SHEATH AND THE SUPPORT CATHETER. PHYSICIAN MADE SEVERAL ATTEMPTS TO SNARE THE WIRE IN THE SHEATH AND EVENTUALLY WAS SUCCESSFUL. CASE WAS THEN ABORTED AFTER THIS LAST, FAILED ATTEMPT. IT TOOK THE PHYSICIAN ROUGHLY 60 MINUTES TO SNARE THE WIRE AND REMOVE IT FROM THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ASAHI ASTATO 30 PERIPHERAL GUIDEWIRE DQX ASAHI INTECC CO., LTD. NA 03065-17051

Patients

Seq Age Sex Outcome Treatment
1 66 YR Hospitalization| R