FDA Adverse Event Malfunction Summary report: N

HOMECHOICE CYCLER-REFURBISHED

MDR report key: 1880271 · Received October 22, 2010

Report

Report Number
1423500-2010-04809
Event Type
Malfunction
Date Received
October 22, 2010
Date of Event
September 29, 2010
Report Date
September 30, 2010
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K053512
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). A REVIEW OF THE PREVIOUS SERVICE RECORD, (B)(6) 2010, SHOWS THE DEVICE PASSED ALL REQUIRED TESTS AND CALIBRATIONS PRIOR TO ITS RELEASE FROM THE (B)(6) FACILITY. NO ISSUES WERE IDENTIFIED THAT MAY HAVE CONTRIBUTED TO THE EARTH LEAKAGE CURRENT FAILED. THE PREVIOUS RETURN OF THE DEVICE WAS NOT FOR ANY ISSUES RELATED TO EARTH LEAKAGE CURRENT FAILED. THE ROOT CAUSE WAS DETERMINED TO BE A HARDWARE PROBLEM- SHORT CIRCUIT IN THE PUMP. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION EXPECTED BUT NOT YET COMPLETED. ANY RESULTS OF EVALUATION WILL BE PROVIDED IN A FOLLOW UP E MDR.

Description of Event or Problem · 1

DURING INITIAL ASSESSMENT OF A RETURNED HOMECHOICE (HC) DEVICE, A BAXTER TECHNICIAN DETERMINED THE HC MACHINE SYSTEM FAILED RETURNED INSTRUMENT TEST/EVALUATION TESTING DUE TO AN EARTH LEAKAGE CURRENT FAILED TO MEET SPECIFICATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE CYCLER-REFURBISHED SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1