UNKNOWN TRUFILL PUSHABLE COIL
Report
- Report Number
- 1058196-2010-00305
- Event Type
- Malfunction
- Date Received
- October 22, 2010
- Date of Event
- September 15, 2010
- Report Date
- September 27, 2010
- Manufacturer
- CORDIS NEUROVASCULAR, INC.
- Product Code
- MCG
- PMA / PMN Number
- K983483
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THE DEVICES ARE NOT AVAILABLE FOR ANALYSIS AND THE LOT NUMBER OF THE TRUFILL COIL IS NOT KNOWN; THEREFORE A DEVICE HISTORY RECORD IS NOT POSSIBLE. PROCEDURAL FACTORS THAT MAY HAVE IMPACTED THE EVENT INCLUDE VESSEL/LESION CHARACTERISTICS, INADEQUATE CONTINUOUS FLUSH THROUGH THE MICROCATHETER, AND DEVICE INTERACTION. BASED ON THE LACK OF CLINICAL INFORMATION AND WITHOUT THE RETURN OF THE DEVICE FOR ANALYSIS, NO CONCLUSION CAN BE MADE REGARDING THE TRUFILL BECOMING IMPEDED IN THE UNKNOWN MICROCATHETER WHEN ADVANCED WITH THE TRUFILL COIL PUSHER. THEREFORE, NO CORRECTIVE ACTIONS WILL BE TAKEN AT THIS TIME. PLEASE NOTE: THE CATALOG AND LOT NUMBER/(S) FOR THE ACTUAL PRODUCT/(S) USED IN THE PROCEDURE ARE UNKNOWN. AN INVESTIGATION IS IN PROCESS TO RETRIEVE THIS INFORMATION.
DURING A COIL EMBOLIZATION OF AN UNKNOWN TARGET SITE WHEN THE TRUFILL COIL (UNKNOWN PRODUCT CODE) WAS ADVANCED USING THE TRUFILL COIL PUSHER THE COIL WAS STUCK AND STOPPED ADVANCING THROUGH THE UNKNOWN MICROCATHETER. THE SYSTEM WAS REMOVED AND CHANGED TO OTHER PRODUCTS. THE PROCEDURE WAS COMPLETED WITHOUT ANY PATIENT INJURY. THERE WERE NO CONTRIBUTING ISSUES NOTED WITH THE COIL PUSHER OR THE MICROCATHETER. THE MICROCATHETER WAS FLUSHED AFTER EVERY COIL WAS DELIVERED. NO ADDITIONAL INFORMATION IS KNOWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNKNOWN TRUFILL PUSHABLE COIL | CNV_PUSHABLE COILS (MCG) | MCG | CORDIS NEUROVASCULAR, INC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN MICROCATHETER |