FDA Adverse Event
Malfunction
Summary report: N
DIMENSION® CLINICAL CHEMISTRY SYSTEM
MDR report key: 1880167
·
Received October 22, 2010
Report
- Report Number
- 1226181-2010-00142
- Event Type
- Malfunction
- Date Received
- October 22, 2010
- Date of Event
- October 17, 2010
- Report Date
- October 18, 2010
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS
- Product Code
- MMI
- PMA / PMN Number
- K944093
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
Additional Manufacturer Narrative · 1
ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY ELEVATED TROPONIN I RESULT WAS A MALFUNCTION OF THE HM WASH PUMP CASSETTES. THE CUSTOMER PERFORMED RECOMMENDED MAINTENANCE. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
Description of Event or Problem · 1
A FALSELY ELEVATED TROPONIN I RESULT WAS OBTAINED ON A PATIENT SAMPLE. THE RESULT WAS REPORTED TO THE PHYSICIAN WHO QUESTIONED THE RESULT. THE SAMPLE WAS REPEATED AND A NEGATIVE RESULT WAS OBTAINED. PATIENT TREATMENT WAS NOT ALTERED OR PRESCRIBED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY ELEVATED TROPONIN I RESULT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DIMENSION® CLINICAL CHEMISTRY SYSTEM | CLINICAL CHEMISTRY SYSTEM | MMI | SIEMENS HEALTHCARE DIAGNOSTICS | RXLMAXRH |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |