INTEGRATED APD SET W/CASSETTE3 PRONG
Report
- Report Number
- 1423500-2010-04777
- Event Type
- Malfunction
- Date Received
- October 22, 2010
- Date of Event
- September 30, 2010
- Report Date
- September 30, 2010
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE SAMPLE WAS DISCARDED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW UP MDR WILL BE SENT.
(B)(4). THIS COMPLAINT FOR A SYSTEM ERROR 2240 (AIR IN SET) OCCURRED DURING DRAIN 4/4 WAS NOT CONFIRMED DUE TO A LACK OF SAMPLE; HOWEVER, PER THE COMPLAINT INFORMATION THE CAUSE OF THE SE 2240 IS DUE TO AIR BEING SUCKED INTO THE DISPOSABLE AFTER THE PATIENT DISCONNECTED THE PATIENT LINE FROM THE TRANSFER SET. THE PATIENT DISCONNECTED DURING THE DWELL CYCLE AND RECONNECTED BEFORE THE HC ALARMED SYSTEM ERROR 2240. THIS REVIEW FINDS THE LABELING ADEQUATE FOR THE USE ERROR(S) IN THE COMPLAINT. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).
A PATIENT CONTACTED GLOBAL TECHNICAL SERVICES (GTS) REGARDING ASSISTANCE WITH A SYSTEM ERROR 2240 THAT APPEARED WHILE USING THE HOMECHOICE (HC) DURING DRAIN 4 OF 4. THE PATIENT DISCONNECTED DURING THE DWELL CYCLE AND RECONNECTED BEFORE THE HC ALARMED SYSTEM ERROR 2240. GTS EXPLAINED THAT A LARGE AMOUNT OF AIR HAD ENTERED INTO THE DISPOSABLE. GTS HELPED THE PATIENT CLEAR THE ERROR BY CYCLING POWER TWICE TO THE "PRESS GO TO START" PROMPT. THE PATIENT AGREED TO CONTACT THEIR DIALYSIS NURSE AND ELECTED TO COMPLETE THERAPY WITH MANUAL SUPPLIES. GTS REVIEWED PROPER PROCEDURE WITH THE PATIENT. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INTEGRATED APD SET W/CASSETTE3 PRONG | SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR |