FDA Adverse Event Malfunction Summary report: N

7900

MDR report key: 1880027 · Received October 13, 2010

Report

Report Number
9680959-2010-00396
Event Type
Malfunction
Date Received
October 13, 2010
Date of Event
August 17, 2010
Report Date
October 13, 2010
Manufacturer
GE OEC MEDICAL SYSTEMS GMBH
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE F4 FUSE WAS REPLACED DURING THE SERVICE CALL.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE 7900 SYSTEM WOULD NOT BOOT UP. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 7900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS GMBH 7900

Patients

Seq Age Sex Outcome Treatment
1