FDA Adverse Event Malfunction Summary report: N

BARCO

MDR report key: 18796478 · Received February 28, 2024

Report

Report Number
3004455125-2024-00012
Event Type
Malfunction
Date Received
February 28, 2024
Date of Event
June 9, 2023
Report Date
February 14, 2024
Manufacturer
BARCO NV
Product Code
DXJ
PMA / PMN Number
K173381
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

BARCO HAS REQUESTED TO RECEIVE MORE INFORMATION FROM THE CUSTOMER, HOWEVER AFTER SEVERAL ATTEMPTS THE CUSTOMER DID NOT PROVIDE ANY FEEDBACK.

Description of Event or Problem · 0

REPORTED FAILURE: DURING AN OPERATION, THE IMAGE FROM THE CONNECTED ENDOSCOPE DISAPPEARED ON ALL 3 DISPLAYS TO WHICH THE SIGNAL WAS ROUTED. WHEN WE CONNECTED TO THE SYSTEM REMOTELY, THE IMAGE WAS THERE AGAIN. WHEN TROUBLESHOOTING, IT TURNED OUT THAT THE CPU UTILIZATION ON THE PC WAS 100% ON ALL 4 CORES. THE OPENJDK PLATFORM BINARY PROCESS WAS CAUSING THE HIGH CPU LOAD. AFTER SHUTTING DOWN AND RESTARTING THE SYSTEM, THE CPU LOAD RETURNED TO NORMAL. TODAY WE RECEIVED ANOTHER TICKET FROM THE OR THAT A DISPLAY DIDN'T SHOW A PICTURE. EVEN NOW THE CPU LOAD WAS AGAIN AT 100%.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
451313 BARCO NEXXISOR, NMS DXJ BARCO NV NEXXISOR NMS

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown